Literature DB >> 22841903

Efficacy of two commercially available cardiac pacing catheters for transesophageal atrial pacing in dogs.

Emily H Chapel1, Robert A Sanders.   

Abstract

OBJECTIVE: Compare the efficacy of two cardiac pacing catheters for transesophageal atrial pacing (TAP). ANIMALS: Ten healthy dogs.
METHODS: Transesophageal atrial pacing was attempted in left lateral recumbency under general anesthesia. In series, a curved electrophysiologic (EP) catheter and a straight transesophageal pacing (TP) catheter were passed transorally into the distal esophagus, caudal to the heart. Each catheter was slowly withdrawn until pacing was initiated. The catheter continued to be withdrawn in 5 mm increments until capture was no longer achieved at maximal pacing setting creating a total zone of capture (tZOC). Minimum pacing threshold (Th(min)) and degree of extraneous muscle stimulation (EMS) were determined at each site.
RESULTS: The EP catheter achieved TAP in 10/10 dogs with a Th(min) of 8.8 ± 4.8 mA and tZOC of 4.0 ± 1.7 cm. The TP catheter achieved TAP in 7/10 dogs with a Th(min) of 22.5 ± 5.6 mA and tZOC of 1.2 ± 1.6 cm. No EMS was noted during TAP at Th(min) using the EP catheter. Of the 7 dogs in which TAP was successful, 3/7 experienced considerable EMS at Th(min). There was no significant relationship between patient body weight and Th(min) with either catheter (EP catheter P = 0.09; TP catheter P = 0.81).
CONCLUSIONS: The electrophysiology catheter captured the atria of more dogs over a larger area without EMS as compared to the TP catheter. The electrophysiology catheter is recommended for transesophageal atrial pacing in dogs.
Copyright © 2012 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22841903      PMCID: PMC3428425          DOI: 10.1016/j.jvc.2012.05.003

Source DB:  PubMed          Journal:  J Vet Cardiol        ISSN: 1760-2734            Impact factor:   1.701


  17 in total

1.  Use of transesophageal atrial pacing to provide temporary chronotropic support in a dog undergoing permanent pacemaker implantation.

Authors:  Robert A Sanders; Henry W Green; Daniel F Hogan; Kim Sederquist
Journal:  J Vet Cardiol       Date:  2011-08-02       Impact factor: 1.701

2.  Decreasing electrode size and increasing efficiency of cardiac stimulation.

Authors:  S Furman; B Parker; D J Escher
Journal:  J Surg Res       Date:  1971-03       Impact factor: 2.192

3.  Low-current bedside emergency atrial and ventricular cardiac pacing from the oesophagus.

Authors:  P Touborg; H R Andersen; P Pless
Journal:  Lancet       Date:  1982-01-16       Impact factor: 79.321

4.  Evaluation of transesophageal atrial pacing in the prone and lateral positions.

Authors:  N M Schwann; D P Maguire; J V Roth; S E McNulty; A Saouaf
Journal:  J Cardiothorac Vasc Anesth       Date:  2001-04       Impact factor: 2.628

5.  Esophageal pacing: a diagnostic and therapeutic tool.

Authors:  J J Gallagher; W M Smith; C R Kerr; J Kasell; L Cook; M Reiter; R Sterba; M Harte
Journal:  Circulation       Date:  1982-02       Impact factor: 29.690

Review 6.  Stimulation and sensing thresholds for cardiac pacing: electrophysiologic and technical aspects.

Authors:  S S Barold; L S Ong; R A Heinle
Journal:  Prog Cardiovasc Dis       Date:  1981 Jul-Aug       Impact factor: 8.194

7.  Transesophageal pacing for prognostic evaluation of preexcitation syndrome and assessment of protective therapy.

Authors:  G Critelli; G Grassi; F Perticone; F Coltorti; V Monda; M Condorelli
Journal:  Am J Cardiol       Date:  1983-02       Impact factor: 2.778

8.  Transesophageal atrial pacing threshold: role of interelectrode spacing, pulse width and catheter insertion depth.

Authors:  D W Benson; M Sanford; A Dunnigan; D G Benditt
Journal:  Am J Cardiol       Date:  1984-01-01       Impact factor: 2.778

9.  Atrial pacing from the esophagus in the diagnosis and management of tachycardia and palpitations.

Authors:  D W Benson; A Dunnigan; R Sterba; D G Benditt
Journal:  J Pediatr       Date:  1983-01       Impact factor: 4.406

10.  The induction of atrial flutter and fibrillation and the termination of atrial flutter by esophageal pacing.

Authors:  C R Kerr; J J Gallagher; W M Smith; R Sterba; L D German; L Cook; J H Kasell
Journal:  Pacing Clin Electrophysiol       Date:  1983-01       Impact factor: 1.976

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.