Pivotal results for the Medtronic Valiant Thoracic Stent Graft System in the VALOR II trial.

Ronald M Fairman1, J Michael Tuchek, W Anthony Lee, Karthikeshwar Kasirajan, Rodney White, Manish Mehta, Sean Lyden, Dipankar Mukherjee, Joseph Bavaria.   

Abstract

OBJECTIVE: We report 30-day and 12-month results of endovascular treatment with the Valiant Thoracic Stent Graft System (Medtronic Vascular, Santa Rosa, Calif) in patients with descending thoracic aortic aneurysms of degenerative etiology. The Valiant stent graft is an evolution of the Talent thoracic stent graft (Medtronic Vascular).
METHODS: The VALOR II (Evaluation of the Clinical Performance of the Valiant Thoracic Stent Graft System in the Treatment of Descending Thoracic Aneurysms of Degenerative Etiology in Subjects Who Are Candidates for Endovascular Repair) was a prospective, nonrandomized, pivotal trial conducted at 24 U.S. sites with enrollment between December 2006 and September 2009. Standard follow-up examinations, including physical examination, computed tomography, and chest radiography, were at 1, 6, and 12 months, and annually through 5 years. VALOR II outcomes were compared with those from the pivotal VALOR (Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms) trial of the Talent stent graft, which enrolled 195 patients with similar enrollment criteria.
RESULTS: VALOR II enrolled 160 patients. Compared with VALOR patients, VALOR II patients had similar age and sex distribution but higher rates of cardiovascular risk factors and significantly more severe modified Society for Vascular Surgery/American Association for Vascular Surgery risk scores. Stent graft delivery and deployment were successful in 154 patients (96.3%). Outcomes at 30 days in VALOR II were perioperative mortality, 3.1%; major adverse events, 38.1%; paraplegia, 0.6%; paraparesis, 1.9%; and stroke, 2.5%. At 12 months, after the minimum sample size was reached, 151 patients were evaluated: aneurysm-related mortality was 4.0%, stent graft migration was 2.9%, and endoleak was 13.0%. Through 12 months, there were no ruptures, conversions to open surgery, secondary procedures due to endoleak >30 days, or loss of stent graft patency. The Valiant stent graft was statistically noninferior to the Talent stent graft in 12-month all-cause mortality (12.6% vs 16.1%) and exceeded the primary effectiveness goal of 12-month successful aneurysm treatment, defined as absence of aneurysm growth >5 mm and of secondary procedures for type I/III endoleak (97.4% vs 80.0%).
CONCLUSIONS: The VALOR II 12-month results demonstrate that the Medtronic Valiant thoracic stent graft is a safe and effective treatment for patients with descending thoracic aortic aneurysms of degenerative etiology.
Copyright © 2012 Society for Vascular Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22832267     DOI: 10.1016/j.jvs.2012.04.062

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  8 in total

1.  Contemporary evaluation of mortality and stroke risk after thoracic endovascular aortic repair.

Authors:  Frances Y Hu; Zachary B Fang; Bradley G Leshnower; Yazan Duwayri; William D Jordan; Theresa W Gillespie; Ravi K Veeraswamy
Journal:  J Vasc Surg       Date:  2017-05-11       Impact factor: 4.268

2.  Long-term results of endovascular repair for distal arch and descending thoracic aortic aneurysms treated by custom-made endografts: usefulness of fenestrated endografts.

Authors:  Masakazu Matsuyama; Kunihide Nakamura; Hiroyuki Nagahama; Katsuhiko Nina; Jouji Endou; Kazushi Kojima; Masanori Nishimura; Hirohito Ishii; Atsuko Yokota
Journal:  Ann Vasc Dis       Date:  2014-12-25

Review 3.  Open Versus Endovascular or Hybrid Thoracic Aortic Aneurysm Repair.

Authors:  Ryan Clare; Julianne Jorgensen; Somjot S Brar
Journal:  Curr Atheroscler Rep       Date:  2016-10       Impact factor: 5.113

Review 4.  Endovascular Repair in Acute Complicated Type B Aortic Dissection: 3-Year Results from the Valiant US Investigational Device Exemption Study.

Authors:  Chang Young Lim
Journal:  Korean J Thorac Cardiovasc Surg       Date:  2017-06-05

5.  Thoracic Endovascular Aortic Repair Adverse Events Reported In The Food And Drug Administration Manufacturer And User Facility Device Experience Database.

Authors:  Neel A Mansukhani; Meraaj S Haleem; Mark K Eskandari
Journal:  Med Devices (Auckl)       Date:  2019-11-14

6.  Clinical features and prognosis of patients with and without diabetes mellitus undergoing endovascular aortic aneurysm repair.

Authors:  Mitsuyoshi Takahara; Osamu Iida; Junichi Tazaki; Ryusuke Nishikawa; Kiyonori Nanto; Yoshiro Chiba; Kazuhisa Sakamoto; Makoto Kinoshita; Naoki Takahashi; Satoshi Kamihira; Terutoshi Yamaoka; Hirooki Higami; Takeichiro Nakane; Takahiro Ohmine; Atsushi Guntani
Journal:  BMC Endocr Disord       Date:  2022-04-07       Impact factor: 2.763

7.  Research and clinical translation of trilayer stent-graft of expanded polytetrafluoroethylene for interventional treatment of aortic dissection.

Authors:  Gang Wang; Caiyun Gao; Benhao Xiao; Jie Zhang; Xunyuan Jiang; Qunsong Wang; Jingzhen Guo; Deyuan Zhang; Jianxiong Liu; Yuehui Xie; Chang Shu; Jiandong Ding
Journal:  Regen Biomater       Date:  2022-07-22

Review 8.  Secondary Endoleak Management Following TEVAR and EVAR.

Authors:  Seyed Ameli-Renani; Vyzantios Pavlidis; Robert A Morgan
Journal:  Cardiovasc Intervent Radiol       Date:  2020-08-10       Impact factor: 2.740