Comparison of action levels for patient-specific quality assurance of intensity modulated radiation therapy and volumetric modulated arc therapy treatments.

Gordon M Mancuso1, Jonas D Fontenot, John P Gibbons, Brent C Parker.   

Abstract

PURPOSE: To perform a comprehensive and systematic comparison of fixed-beam IMRT and volumetric modulated arc therapy (VMAT) patient-specific QA measurements for a common set of geometries using typical measurement methods.
METHODS: Fixed-beam IMRT and VMAT plans were constructed for structure set geometries provided by AAPM Task Group 119. The plans were repeatedly delivered across multiple measurement sessions, and the resulting dose distributions were measured with (1) radiochromic film and ionization chamber and (2) a commercial two-dimensional diode array. The resulting QA measurements from each delivery technique were then analyzed, compared, and tested for statistically significant differences.
RESULTS: Although differences were noted between QA results for some plans, neither modality showed consistently better agreement of measured and planned doses: of the 22 comparisons, IMRT showed better QA results in 11 cases, and VMAT showed better QA results in 11 cases. No statistically significant differences (p < 0.05) between IMRT and VMAT QA results were found for point doses measured with an ionization chamber, planar doses measured with radiochromic film, or planar doses measured with a two-dimensional diode array.
CONCLUSIONS: These results suggest that it is appropriate to apply patient-specific QA action levels derived from fixed-beam IMRT to VMAT.

Entities:  

Mesh:

Year:  2012        PMID: 22830770     DOI: 10.1118/1.4729738

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  Transitioning from measurement-based to combined patient-specific quality assurance for intensity-modulated proton therapy.

Authors:  Mei Chen; Pablo Yepes; Yoshifumi Hojo; Falk Poenisch; Yupeng Li; Jiayi Chen; Cheng Xu; Xiaodong He; G Brandon Gunn; Steven J Frank; Narayan Sahoo; Heng Li; Xiaorong Ronald Zhu; Xiaodong Zhang
Journal:  Br J Radiol       Date:  2019-12-16       Impact factor: 3.039

2.  Comparison between an in-house 1D profile correction method and a 2D correction provided in Varian's PDPC Package for improving the accuracy of portal dosimetry images.

Authors:  Maritza A Hobson; Stephen D Davis
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

3.  Gamma analysis dependence on specified low-dose thresholds for VMAT QA.

Authors:  Ji-Hye Song; Min-Joo Kim; So-Hyun Park; Seu-Ran Lee; Min-Young Lee; Dong Soo Lee; Tae Suk Suh
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

4.  Investigation on the effect of sharp phantom edges on point dose measurement during patient-specific dosimetry with Rapid Arc.

Authors:  R A Kinhikar; V P Pandey; Rojas K Jose; U Mahantshetty; D S Dhote; D D Deshpande; S K Shrivastava
Journal:  J Med Phys       Date:  2013-07

5.  Reproducibility in patient-specific IMRT QA.

Authors:  Elizabeth M McKenzie; Peter A Balter; Francesco C Stingo; Jimmy Jones; David S Followill; Stephen F Kry
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

6.  IMRT and RapidArc commissioning of a TrueBeam linear accelerator using TG-119 protocol cases.

Authors:  Ning Wen; Bo Zhao; Jinkoo Kim; Karen Chin-Snyder; Maria Bellon; Carri Glide-Hurst; Kenneth Barton; Daiquan Chen; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2014-09-08       Impact factor: 2.102

7.  Correlation of phantom-based and log file patient-specific QA with complexity scores for VMAT.

Authors:  Christina E Agnew; Denise M Irvine; Conor K McGarry
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

8.  An improvement in IMRT QA results and beam matching in linacs using statistical process control.

Authors:  Justin D Gagneur; Gary A Ezzell
Journal:  J Appl Clin Med Phys       Date:  2014-09-08       Impact factor: 2.102

9.  Evaluation of a novel secondary check tool for intensity-modulated radiotherapy treatment planning.

Authors:  Jonas D Fontenot
Journal:  J Appl Clin Med Phys       Date:  2014-09-08       Impact factor: 2.102

10.  Correlation analysis between 2D and quasi-3D gamma evaluations for both intensity-modulated radiation therapy and volumetric modulated arc therapy.

Authors:  Jung-In Kim; Chang Heon Choi; Hong-Gyun Wu; Jin Ho Kim; Kyubo Kim; Jong Min Park
Journal:  Oncotarget       Date:  2017-01-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.