Technical note: patient-specific quality assurance methods for TomoDirect(TM) whole breast treatment delivery.

P Catuzzo1, F Zenone, S Aimonetto, A Peruzzo, V Casanova Borca, M Pasquino, P Franco, M R La Porta, U Ricardi, S Tofani.   

Abstract

PURPOSE: To investigate the feasibility of implementing a novel approach for patient-specific QA of TomoDirect(TM) whole breast treatment.
METHODS: The most currently used TomoTherapy DQA method, consisting in the verification of the 2D dose distribution in a coronal or sagittal plane of the Cheese Phantom by means of gafchromic films, was compared with an alternative approach based on the use of two commercially available diode arrays, MapCHECK2(TM) and ArcCHECK(TM). The TomoDirect(TM) plans of twenty patients with a primary unilateral breast cancer were applied to a CT scan of the Cheese Phantom and a MVCT dataset of the diode arrays. Then measurements of 2D dose distribution were performed and compared with the calculated ones using the gamma analysis method with different sets of DTA and DD criteria (3%-3 mm, 3%-2 mm). The sensitivity of the diode arrays to detect delivery and setup errors was also investigated.
RESULTS: The measured dose distributions showed excellent agreement with the TPS calculations for each detector, with averaged fractions of passed Γ values greater than 95%. The percentage of points satisfying the constraint Γ < 1 was significantly higher for MapCHECK2(TM) than for ArcCHECK(TM) and gafchromic films using both the 3%-3 mm and 3%-2 mm gamma criteria. Both the diode arrays show a good sensitivity to delivery and setup errors using a 3%-2 mm gamma criteria.
CONCLUSIONS: MapCHECK2™ and ArcCHECK(TM) may fulfill the demands of an adequate system for TomoDirect(TM) patient-specific QA.

Entities:  

Mesh:

Year:  2012        PMID: 22830740     DOI: 10.1118/1.4722967

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Intensity-modulated and hypofractionated simultaneous integrated boost adjuvant breast radiation employing statics ports of tomotherapy (TomoDirect): a prospective phase II trial.

Authors:  Pierfrancesco Franco; Michele Zeverino; Fernanda Migliaccio; Domenico Cante; Piera Sciacero; Valeria Casanova Borca; Paolo Torielli; Cecilia Arrichiello; Giuseppe Girelli; Maria Rosa La Porta; Santi Tofani; Gianmauro Numico; Umberto Ricardi
Journal:  J Cancer Res Clin Oncol       Date:  2013-11-30       Impact factor: 4.553

2.  Intensity-modulated adjuvant whole breast radiation delivered with static angle tomotherapy (TomoDirect): a prospective case series.

Authors:  Pierfrancesco Franco; Michele Zeverino; Fernanda Migliaccio; Piera Sciacero; Domenico Cante; Valeria Casanova Borca; Paolo Torielli; Cecilia Arrichiello; Giuseppe Girelli; Gianmauro Numico; Maria Rosa La Porta; Santi Tofani; Umberto Ricardi
Journal:  J Cancer Res Clin Oncol       Date:  2013-09-14       Impact factor: 4.553

3.  The sensitivity of ArcCHECK-based gamma analysis to manufactured errors in helical tomotherapy radiation delivery.

Authors:  Alistair K Templeton; James C H Chu; Julius V Turian
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

4.  Dosimetric comparison of TomoDirect, helical tomotherapy, and volumetric modulated arc therapy for postmastectomy treatment.

Authors:  Wannapha Nobnop; Panchalee Phakoetsuk; Imjai Chitapanarux; Damrongsak Tippanya; Darat Khamchompoo
Journal:  J Appl Clin Med Phys       Date:  2020-07-27       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.