Literature DB >> 22825904

Difficult dual stage transcutaneous lead extraction complicated by fracture of both leads.

Andrzej Kutarski1, Radosław Pietura, Marek Czajkowski.   

Abstract

Extraction of two long-implanted (11 year-old) active leads due to local infection was complicated by the fracture of both leads with a metal sheath. The presence of a temporary pacing lead, inserted due to pacemaker dependence, made extraction even more difficult. The fixation of both proximal extracted lead endings in the occluded subclavian vein was much stronger than the connection of the tips of the leads with the RAA and RVA endocardium. The disintegration of both leads caused by drawing upon the distal ending added additional technical complications to the final procedure. The presented case underscores the necessity of having a full set of tools to hand even during what appears to be a routine lead extraction procedure. The importance of proper lead implantation on its extraction (in what may be the distant future) is discussed.

Entities:  

Mesh:

Year:  2012        PMID: 22825904     DOI: 10.5603/cj.2012.0074

Source DB:  PubMed          Journal:  Cardiol J        ISSN: 1898-018X            Impact factor:   2.737


  3 in total

1.  Echocardiographic assessment of residuals after transvenous intracardiac lead extraction.

Authors:  Magdalena Poterała; Andrzej Kutarski; Wojciech Brzozowski; Michał Tomaszewski; Leszek Gromadziński; Andrzej Tomaszewski
Journal:  Int J Cardiovasc Imaging       Date:  2019-11-16       Impact factor: 2.357

2.  A comprehensive multiscale framework for simulating optogenetics in the heart.

Authors:  Patrick M Boyle; John C Williams; Christina M Ambrosi; Emilia Entcheva; Natalia A Trayanova
Journal:  Nat Commun       Date:  2013       Impact factor: 14.919

3.  Proximal end of 15-year-old ventricular electrode penetrating pulmonary tissue - a source of infection and a challenge for transvenous lead extraction.

Authors:  Maciej Polewczyk; Aneta M Polewczyk; Andrzej Kutarski; Anna Polewczyk
Journal:  Postepy Kardiol Interwencyjnej       Date:  2015-09-28       Impact factor: 1.426

  3 in total