Guidance for storing blood samples in laboratories performing complete blood count with differential.

E Cornet1,2, C Behier1, X Troussard1,2.   

Abstract

INTRODUCTION: The complete blood count (CBC) with differential leukocyte count (DIFF) is an important part of clinical laboratory analyses and provides crucial data for clinicians. Delivery time after blood collection and conditions of storage is known to affect the reliability of results of some hematologic parameters. The aim of this study was to assess the variations of hematologic parameters over time and the influence of storage temperature.
METHODS: Blood samples were randomly selected from hospitalized patients and stored at room temperature and at 4 °C. CBC and DIFF were performed on an automated hematology analyzer and the results between the two groups were compared.
RESULTS: Samples stored at room temperature showed an important increase in mean corpuscular volume and hematocrit and a decrease in mean corpuscular hemoglobin concentration. Neutrophil counts tended to increase, whereas monocyte counts tended to decrease.
CONCLUSION: Storing samples at 4 °C improved reproducibility over time of all quantitative and qualitative parameters. We also observed that NEUT-X, a routine parameter useful in detecting myelodysplastic syndrome, became unreliable when analyzed 24 h after sample collection. Our results led us to recommend that samples should be analyzed within 6 h, particularly if samples are transported at room temperature. We also recommend storing samples at 4 °C in case of remote CBC analysis, especially in the context of clinical trials.
© 2012 Blackwell Publishing Ltd.

Entities:  

Keywords:  Blood cell count; blood preservation; laboratories/standards; reproducibility of results; temperature

Mesh:

Substances:

Year:  2012        PMID: 22823600     DOI: 10.1111/j.1751-553X.2012.01452.x

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  3 in total

1.  Synovial Fluid Interleukin Levels Cannot Distinguish between Prosthetic Joint Infection and Active Rheumatoid Arthritis after Hip or Knee Arthroplasty.

Authors:  Leilei Qin; Chengcheng Du; Jianye Yang; Hai Wang; Xudong Su; Li Wei; Chen Zhao; Cheng Chen; Hong Chen; Ning Hu; Wei Huang
Journal:  Diagnostics (Basel)       Date:  2022-05-11

2.  Comparison of hematologic measurements between local and central laboratories: data from the BABY HUG trial.

Authors:  Ram Kalpatthi; Bruce Thompson; Ming Lu; Winfred C Wang; Niren Patel; Abdullah Kutlar; Thomas Howard; Lori Luchtman-Jones; Scott T Miller
Journal:  Clin Biochem       Date:  2012-10-31       Impact factor: 3.281

3.  The Copenhagen Primary Care Differential Count (CopDiff) database.

Authors:  Christen Lykkegaard Andersen; Volkert Dirk Siersma; Willy Karlslund; Hans Carl Hasselbalch; Peter Felding; Ole Weis Bjerrum; Niels de Fine Olivarius
Journal:  Clin Epidemiol       Date:  2014-06-12       Impact factor: 4.790

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.