Literature DB >> 22823449

Safety and long-term outcome of modified intracardiac echocardiography-assisted "no-balloon" sizing technique for transcatheter closure of ostium secundum atrial septal defect.

Gianluca Rigatelli1, Fabio Dell'avvocata, Paolo Cardaioli, Massimo Giordan, Ho Thuong Dung, Nguyen Thuong Nghia, Ramesh Daggubati, Aravinda Nanjiundappa.   

Abstract

BACKGROUND: The need for sizing the secundum atrial septal defect (ASD) with the balloon sizing technique is still debated at least in adult patients. We sought to prospectively evaluate the effectiveness of intracardiac echocardiography (ICE)-aided sizing technique for transcatheter closure of secundum ASD, without using a balloon sizing.
METHODS: In a prospective 5-year registry, we enrolled 81 patients (mean age 48 ± 13.7 years, 54 females) who had been referred to three different centers for atheter-based closure of secundum ASD. Eligible patients underwent ICE study and closure attempt. In a preliminary group of 21 patients, sizing balloon was performed under ICE guidance to assess the value of rim thickness necessary for device anchorage. In the remaining 60 patients, the retrieved value of the rim thickness was measured on ICE and used as key points to measure the defect and select the device.
RESULTS: In the preliminary group of patients, the value of thickness at point of initial deflection by the balloon was 1.23 ± 0.1 mm. ASD diameter in the study group was measured at the point of rim with at least 1.2 mm and the mean ASD diameter was 26.2 ± 10.1 mm. Rates of procedural success, predischarge occlusion, and major complications rate were 100%, 93.3%, and 0%, respectively. On mean follow-up of 5.4 ± 1.8 years, the occlusion rate was 98.7% with no long-term complications.
CONCLUSIONS: Our novel ICE-sizing technique appears to be safe and effective in adult patients, thus eventually minimizing overestimation, costs, and potential complications of balloon sizing. ©2012, Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22823449     DOI: 10.1111/j.1540-8183.2012.00755.x

Source DB:  PubMed          Journal:  J Interv Cardiol        ISSN: 0896-4327            Impact factor:   2.279


  4 in total

1.  Small left atrial size complicating percutaneous transcatheter device closure of secundum atrial septal defect with conventional approach.

Authors:  Hong Ki Ko; So Yeon Kang; Jeong Jin Yu; Jae-Kon Ko; Young-Hwue Kim
Journal:  Korean Circ J       Date:  2015-05-27       Impact factor: 3.243

Review 2.  Comprehensive understanding of atrial septal defects by imaging studies for successful transcatheter closure.

Authors:  Jinyoung Song
Journal:  Korean J Pediatr       Date:  2014-07-23

3.  Phase contrast cardiac magnetic resonance imaging versus transoesophageal echocardiography for the evaluation of feasibility for transcatheter closure of atrial septal defects.

Authors:  Tariq Rashid Shah; Jahangir Rashid Beig; Naseer Ahmad Choh; Fayaz Ahmad Rather; Irfan Yaqoob; Vicar Mohammad Jan
Journal:  Egypt Heart J       Date:  2022-04-13

4.  Trends in the Prevalence of Atrial Septal Defect and Its Associated Factors among Congenital Heart Disease Patients in Vietnam.

Authors:  Ho Xuan Tuan; Phan The Phuoc Long; Vu Duy Kien; Le Manh Cuong; Nguyen Van Son; Robert Dalla-Pozza
Journal:  J Cardiovasc Dev Dis       Date:  2019-12-27
  4 in total