Literature DB >> 22821155

[Influence of high-definition multileaf collimator for three-dimensional conformal radiotherapy and intensity-modulated radiotherapy of prostate cancer].

Sachie Fujimoto1, Kaoru Ono, Kengo Furukawa, Tsuyoshi Kudo, Yukio Akagi, Tadashi Koyama, Yutaka Hirokawa, Shintarou Kawai, Takeo Nakashima.   

Abstract

The focus of this work is to evaluate the dosimetric impact of treatment planning for three-dimensional conformal radiotherapy (3DCRT) and intensity-modulated radiotherapy (IMRT) of prostate cancer using Varian/BrainLAB 120-leaf high-definition multileaf collimator (HD120 MLC) with 2.5 mm leaf width and Varian 120-leaf millennium multileaf collimator (M120 MLC) with 5 mm leaf width. We measured the leaf transmission and dosimetric leaf gap (DLG) of two multileaf collimator (MLC) systems using Farmer ionization chamber. The dosimetric impact of treatment planning for 3DCRT and IMRT of prostate cancer for ten clinical cases using two MLC systems was evaluated quantitatively. 3DCRT was divided to 3DCRT(middle) as fitting at middle of leaf tip and 3DCRT(outside) as fitting at outside of leaf tip. The leaf transmission factor and DLG of HD120 MLC for 6 and 10 MV X-ray decreased by 0.2% and 1 mm, respectively, compared to M120 MLC. The mean conformity index of PTV of treatment planning for prostate 3DCRT(middle), 3DCRT(outside) , and IMRT decreased by 0.9%, 6.6%, and 0.9% and the mean homogeneity index increased 2.3%, 13.0%, and 4.2%, respectively. The mean V20, V40, and V65 decreased by 2.4%, 6.6%, and 4.5% for bladder and 3.3%, 6.1%, and 5.9% for rectum, respectively. The results of this work demonstrated that the dose conformity of PTV improved and the dose of bladder and rectum decreased for 3DCRT and IMRT of prostate cancer using HD120 MLC compared to M120 MLC, because of reduction of leaf width, leaf transmission, and rounded leaf end transmission.

Entities:  

Mesh:

Year:  2012        PMID: 22821155     DOI: 10.6009/jjrt.2012_jsrt_68.7.825

Source DB:  PubMed          Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi        ISSN: 0369-4305


  2 in total

1.  Influence of multi-leaf collimator leaf transmission on head and neck intensity-modulated radiation therapy and volumetric-modulated arc therapy planning.

Authors:  Motoharu Sasaki; Masahide Tominaga; Takeshi Kamomae; Hitoshi Ikushima; Motonori Kitaoka; Ryota Bando; Kanako Sakuragawa; Masataka Oita
Journal:  Jpn J Radiol       Date:  2017-06-24       Impact factor: 2.374

2.  A systematic review and meta-analysis of radiotherapy planning studies comparing multi leaf collimator designs.

Authors:  Marco Luzzara; Luigi Santoro; Kevin Brown
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-23
  2 in total