| Literature DB >> 22816247 |
Donald L Gilliland1, Charles K Black, James E Denison, Charles T Seipelt.
Abstract
During the "Standards Development and International Harmonization: AOAC INTERNATIONAL Mid-Year Meeting" held on June 29, 2011, an Expert Review Panel (ERP) on behalf of AOAC INTERNATIONAL adopted the method "Simultaneous Determination of Vitamins D2 and D3 by LC-MS/MS in Infant Formula and Adult Nutritionals" as an AOAC Official First Action method. Vitamins D2 and D3 are extracted from the sample using pentane-ether; the extract is collected and dried under nitrogen. Vitamin D is separated from interfering compounds using UPLC, and quantitated using tandem mass spectrometry (MS/MS). Preliminary data showed the intermediate precision ranged from 3.34-8.05% and an accuracy range of 98.5-111% over the samples tested for vitamin D3. For vitamin D2, the intermediate precision ranged from 2.37-5.45% and accuracy ranged from 96.4-104% over the four matrixes evaluated. The analytical range for the method is bounded by the concentrations of the working standards, 21-270 ng/mL, and is equivalent to 0.168-2.16 mcg/100 g in ready-to-feed product. The practical method quantitation limit is 0.168 mcg/100 g product with method detection limit of 60 ng/100 g product. The ERP reviewed the data and determined that the performance characteristics of the method met the standard method performance requirements, and therefore the method was granted First Action status.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22816247 DOI: 10.5740/jaoacint.cs2011_13
Source DB: PubMed Journal: J AOAC Int ISSN: 1060-3271 Impact factor: 1.913