Optimal parameters to improve the interface quality of the flap bed in femtosecond laser-assisted laser in situ keratomileusis.

Sebastiano Serrao1, Lucio Buratto, Giuseppe Lombardo, Maria P De Santo, Pietro Ducoli, Marco Lombardo.   

Abstract

PURPOSE: To analyze the interface quality of the anterior stroma after femtosecond laser flap creation using atomic force microscopy.
SETTING: IRCCS Fondazione G.B. Bietti, Rome, Italy.
DESIGN: Experimental study.
METHODS: A 110 μm depth flap was created in 20 human corneal tissues using a femtosecond laser platform (Intralase iFS). Tissues were divided into 4 groups of various cutting parameters: pulse energy and spot separation of 0.75 μJ and 6 μm (Group 1), 0.65 μJ and 5 μm (Group 2), 0.55 μJ and 4 μm (Group 3), and 0.45 μJ and 4 μm (Group 4). Four additional tissue sections were cut using a motorized microkeratome (Hansatome). Atomic force microscopy (Autoprobe CP) analysis was performed on the stromal bed of each sample.
RESULTS: The corneal tissues treated with higher pulse energies and wider spot separations (Groups 1 and 2) showed a rougher stromal bed interface (root mean square [RMS] rough = 0.23 μm ± 0.008 (SD) and 0.24 ± 0.009 μm, respectively) than tissues in Groups 3 and 4 (RMS rough = 0.18 ± 0.006 μm and 0.18 ± 0.008 μm, respectively; P<.001, 1-way analysis of variance). The stromal surface quality of tissues treated with pulse energies of 0.55 μJ or lower and 4 μm spot separation compared favorably with that of tissues cut by the microkeratome (RMS rough = 0.17 ± 0.006 μm; P>.05, Tukey).
CONCLUSIONS: The femtosecond stromal interface quality was improved with pulse energy lower and spot separations narrower than those currently used in the clinical setting. The flap interface smoothness created by the femtosecond laser was comparable to that created by the microkeratome. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2012 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22814052     DOI: 10.1016/j.jcrs.2012.05.021

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  7 in total

1.  A microscopic study of the corneal stromal lenticules extracted during femtosecond laser-assisted small incision lenticule extraction.

Authors:  Yewei Yin; Tu Hu; Aiqun Xiang; Yanyan Fu; Yang Zhao; Xiaoying Wu; Xiaoying Wu; Dan Wen
Journal:  Exp Ther Med       Date:  2021-04-25       Impact factor: 2.447

2.  Interface quality of different corneal lamellar-cut depths for femtosecond laser-assisted lamellar anterior keratoplasty.

Authors:  Chenxing Zhang; Matthew Bald; Maolong Tang; Yan Li; David Huang
Journal:  J Cataract Refract Surg       Date:  2015-03-04       Impact factor: 3.351

3.  Comparison of four methods of surface roughness assessment of corneal stromal bed after lamellar cutting.

Authors:  Clotilde Jumelle; Alina Hamri; Gregory Egaud; Cyril Mauclair; Stephanie Reynaud; Virginie Dumas; Sandrine Pereira; Thibaud Garcin; Philippe Gain; Gilles Thuret
Journal:  Biomed Opt Express       Date:  2017-10-12       Impact factor: 3.732

4.  Analysis of femtosecond laser assisted capsulotomy cutting edges and manual capsulorhexis using environmental scanning electron microscopy.

Authors:  Sebastiano Serrao; Giuseppe Lombardo; Giovanni Desiderio; Lucio Buratto; Domenico Schiano-Lomoriello; Marco Pileri; Marco Lombardo
Journal:  J Ophthalmol       Date:  2014-11-20       Impact factor: 1.909

5.  Comparison of visual quality after EVO-ICL implantation and SMILE to select the appropriate surgical method for high myopia.

Authors:  Qin Qin; Lianyun Bao; Liping Yang; Zifang He; Zhenping Huang
Journal:  BMC Ophthalmol       Date:  2019-02-08       Impact factor: 2.209

6.  Comparison of self-reported quality of vision outcomes after myopic LASIK with two femtosecond lasers: a prospective, eye-to-eye study.

Authors:  Christopher S Sáles; Edward E Manche
Journal:  Clin Ophthalmol       Date:  2016-09-01

7.  Comparison of the optical quality after SMILE and FS-LASIK for high myopia by OQAS and iTrace analyzer: a one-year retrospective study.

Authors:  Yewei Yin; Ying Lu; Aiqun Xiang; Yanyan Fu; Yang Zhao; Yuanjun Li; Tu Hu; Kaixuan Du; Shengfa Hu; Qiuman Fu; Xiaoying Wu; Dan Wen
Journal:  BMC Ophthalmol       Date:  2021-08-02       Impact factor: 2.209

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.