An unusual case of a primary blood collection tube with floating separator gel.

Massimo Daves1, Giuseppe Lippi, Giovanni Cosio, Alberto Raffagnini, Evi Peer, Andrea Dangella, Rainer Oberhollenzer, Roberto Cemin.   

Abstract

We describe an unusual case of a primary evacuated blood collection tube with floating separator gel, which has been collected from a 50-year-old man submitted to a percutaneous coronary intervention (PCI). The sample was collected from the femoral artery in a primary evacuated blood collection tube containing lithium-heparin. After centrifugation of the specimen, an unusual positioning of the separator gel was observed, which migrated at the topmost layer, whereas the packed blood cells remained in the middle and the plasma at the bottom. The potential interfering substance was found to be a contrast dye, 140 ml of which were administered to the patient during a revascularization procedure for acute myocardial infarction. The potential aspiration of the gel inappropriately positioned at the top of the tube by laboratory instrumentation can produce several technical and clinical problems, when not reliably detected. First, the needle of the instrument might be partially or completely obstructed by the gel, thus jeopardizing the integrity and correct functioning of the instrument. The aspiration of gel along with the sample matrix might also spuriously modify the test results, since an inappropriate amount of serum or plasma would be analyzed.
© 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22811356      PMCID: PMC6807415          DOI: 10.1002/jcla.21512

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  6 in total

1.  Iodinated contrast media interfere with gel barrier formation in plasma and serum separator tubes.

Authors:  Tom Spiritus; Zahur Zaman; Walter Desmet
Journal:  Clin Chem       Date:  2003-07       Impact factor: 8.327

2.  High total protein impairs appropriate gel barrier formation in BD Vacutainer blood collection tubes.

Authors:  Johannes M W van den Ouweland; Stephan Church
Journal:  Clin Chem       Date:  2007-02       Impact factor: 8.327

3.  Anomalous flotation of separator gel: density or viscosity?

Authors:  María Fatás; Pablo Franquelo; Ricardo Franquelo
Journal:  Clin Chem       Date:  2008-04       Impact factor: 8.327

Review 4.  Impact of blood collection devices on clinical chemistry assays.

Authors:  Raffick A R Bowen; Glen L Hortin; Gyorgy Csako; Oscar H Otañez; Alan T Remaley
Journal:  Clin Biochem       Date:  2009-10-12       Impact factor: 3.281

5.  Solution densities and estimated total protein contents associated with inappropriate flotation of separator gel in different blood collection tubes.

Authors:  Ronald C Faught; James Marshall; Joshua Bornhorst
Journal:  Arch Pathol Lab Med       Date:  2011-09       Impact factor: 5.534

6.  The case of the floating gel.

Authors:  R Srivastava; M J Murphy; J Card; A Severn; C G Fraser
Journal:  J Clin Pathol       Date:  2004-12       Impact factor: 3.411

  6 in total
  6 in total

Review 1.  Pre-analytic variability in cardiovascular biomarker testing.

Authors:  Roberto Cemin; Massimo Daves
Journal:  J Thorac Dis       Date:  2015-10       Impact factor: 2.895

2.  Abnormal gel flotation caused by contrast media during adrenal vein sampling.

Authors:  Gabriel Lima-Oliveira; Giuseppe Lippi; Gian Luca Salvagno; Matteo Gelati; Antonella Bassi; Alberto Contro; Francesca Pizzolo; Gian Cesare Guidi
Journal:  Biochem Med (Zagreb)       Date:  2016-10-15       Impact factor: 2.313

3.  Abnormal gel flotation in a patient with apperant pneumonia diagnosis: a case report.

Authors:  Fethullah Gerin; Dilber Coban Ramazan; Ozgur Baykan; Onder Sirikci; Goncagul Haklar
Journal:  Biochem Med (Zagreb)       Date:  2014-02-15       Impact factor: 2.313

4.  Evaluation of sample hemolysis in blood collected by S-Monovette using vacuum or aspiration mode.

Authors:  Giuseppe Lippi; Paola Avanzini; Roberta Musa; Franca Sandei; Rosalia Aloe; Gianfranco Cervellin
Journal:  Biochem Med (Zagreb)       Date:  2013       Impact factor: 2.313

Review 5.  Interference of medical contrast media on laboratory testing.

Authors:  Giuseppe Lippi; Massimo Daves; Camilla Mattiuzzi
Journal:  Biochem Med (Zagreb)       Date:  2014-02-15       Impact factor: 2.313

6.  Switching from serum to plasma: Implementation of BD Vacutainer® Barricor™ Plasma Blood Collection Tubes improves sample quality and laboratory turnaround time.

Authors:  Christian Ramakers; Brendan Meyer; Wanfei Yang; Elizabeth Plokhoy; Yan Xiong; Stephen Church; Nitin Kaushik
Journal:  Pract Lab Med       Date:  2019-12-10
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.