Literature DB >> 22810548

Does application position of the T-POD affect stability of pelvic fractures?

Mark L Prasarn1, John Small, Bryan Conrad, Nicole Horodyski, Marybeth Horodyski, Glenn R Rechtine.   

Abstract

OBJECTIVE: Most trauma centers place pelvic binders on unstable pelvic fractures for acute management and control of hemorrhage. It has been proposed that the binders be placed at the level of the greater trochanters of the femur. Our hypothesis was that application of the T-POD at this site would provide better immobilization of an unstable pelvic injury than a more cephalad location.
METHODS: Unstable pelvic injuries (OTA type 61-C1) were surgically created in 9 fresh whole human cadavers. Electromagnetic sensors were affixed to the intact and injured sides of the pelvis. A Fastrak, three-dimensional electromagnetic motion analysis device was used to determine the angular motion occurring at the fractured sites. Maximum displacements for sagittal, coronal, and axial rotation were recorded during application of the binder, while performing bed transfers, while logrolling, and elevating the head of the bed. The T-POD device was placed either over the greater trochanters or at the level of the anterior superior iliac spine as per manufacturer's recommendations.
RESULTS: There were no significant differences in the amount of motion produced during application of the T-POD at either location. There was less motion observed in all planes of motion during all maneuvers when the T-POD was placed at the level of the greater trochanters versus anterior superior iliac spine. During bed transfers, this was statistically significant in all planes. This was statistically significant while logrolling in the axial plane and the coronal plane during head of bed elevation.
CONCLUSIONS: We advocate the placement of pelvic binder devices at the level of the greater trochanters for improved control of the fracture in an unstable pelvic injury. This may result in improved control of hemorrhage, better access to the abdomen, and greater patient comfort.

Entities:  

Mesh:

Year:  2013        PMID: 22810548     DOI: 10.1097/BOT.0b013e31826913d6

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  8 in total

1.  An assessment of pelvic binder placement at a UK major trauma centre.

Authors:  H Naseem; P D Nesbitt; D C Sprott; A Clayson
Journal:  Ann R Coll Surg Engl       Date:  2017-09-15       Impact factor: 1.891

2.  Resuscitation of Polytrauma Patients: The Management of Massive Skeletal Bleeding.

Authors:  Enrique Guerado; Maria Luisa Bertrand; Luis Valdes; Encarnacion Cruz; Juan Ramon Cano
Journal:  Open Orthop J       Date:  2015-07-31

Review 3.  Effectiveness of non invasive external pelvic compression: a systematic review of the literature.

Authors:  Peyman Bakhshayesh; Tarek Boutefnouchet; Anna Tötterman
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2016-05-18       Impact factor: 2.953

4.  The T-pod is as stable as supraacetabular fixation using 1 or 2 Schanz screws in partially unstable pelvic fractures: a biomechanical study.

Authors:  Christian Zeckey; Adrian Cavalcanti Kußmaul; Eduardo M Suero; Christian Kammerlander; Axel Greiner; Matthias Woiczinski; Christian Braun; Wilhelm Flatz; Wolfgang Boecker; Christopher A Becker
Journal:  Eur J Med Res       Date:  2020-07-18       Impact factor: 2.175

5.  Experience and Availability of Pelvic Binders at Swedish Trauma Units; A Nationwide Survey.

Authors:  Peyman Bakhshayesh; Sara Heljesten; Lars Weidenhielm; Anders Enocson
Journal:  Bull Emerg Trauma       Date:  2018-07

6.  Three Dimensional Quality Assessments of Applied Pelvic Binders.

Authors:  Peyman Bakhshayesh; David Hullberg Risling; Anders Enocson
Journal:  Bull Emerg Trauma       Date:  2019-04

7.  Does the accuracy of prehospital pelvic binder placement affect cardiovascular physiological parameters during rescue? A clinical study in patients with pelvic ring injuries.

Authors:  Simon Tiziani; Anna-Staschia Janett; Hatem Alkadhi; Georg Osterhoff; Kai Sprengel; Hans-Christoph Pape
Journal:  OTA Int       Date:  2022-04-18

8.  Damage control orthopaedics: State of the art.

Authors:  Enrique Guerado; Maria Luisa Bertrand; Juan Ramon Cano; Ana María Cerván; Adolfo Galán
Journal:  World J Orthop       Date:  2019-01-18
  8 in total