Literature DB >> 22806510

The unacknowledged incidence of laparoscopic stapler malfunction.

Douglas Kwazneski1, Cheryl Six, Kurt Stahlfeld.   

Abstract

BACKGROUND: Laparoscopic instruments are being used with increasing frequency. Our surgeons recently experienced several independent adverse events involving the laparoscopic linear stapler. Although the Food and Drug Administration maintains a Manufacturer and User Facility Device Experience (MAUDE) database to track such voluntary reports, many events are not reported and the true incidence of adverse events is unknown. We attempted to determine how frequently minimally invasive surgeons have experienced technical problems with a laparoscopic stapler.
METHODS: Following IRB approval, we electronically distributed an anonymous 10-question survey to the 124 minimally invasive program directors listed in the Fellowship Council database. The questions focused on personal or peer experience with laparoscopic stapler malfunction, frequency and type of malfunction, device manufacturer, whether the operation was altered, and root cause analysis of the event.
RESULTS: Forty-four of the 124 program directors (35%) completed the survey. The majority reported personal or peer experience (86%) with a linear stapler not releasing (66%) or not firing (73%) after application, with 27% of the respondents noting that this occurred three or more times. The malfunction was not related to type of load, straight (23%) or reticulating (32%) model, or manufacturer (Ethicon 30%, Covidien 36%). One quarter of the respondents noted that the malfunction caused them to significantly alter their operative procedure, and 30% reported that they received no helpful feedback from the manufacturer despite contacting it.
CONCLUSIONS: Most minimally invasive surgeons have experienced laparoscopic linear stapler malfunction and 25% have had to significantly alter the planned operative procedure due to the malfunction.

Mesh:

Year:  2012        PMID: 22806510     DOI: 10.1007/s00464-012-2417-y

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  3 in total

1.  Surgical stapler-associated fatalities and adverse events reported to the Food and Drug Administration.

Authors:  S Lori Brown; Eileen K Woo
Journal:  J Am Coll Surg       Date:  2004-09       Impact factor: 6.113

Review 2.  An unusual intraoperative complication of stapled colorectal anastomosis: report of a case and review of the literature.

Authors:  V Bard; Z Cohen; T Ross
Journal:  Can J Surg       Date:  1993-04       Impact factor: 2.089

3.  Prospective randomized comparison of linear staplers during laparoscopic Roux-en-Y gastric bypass.

Authors:  J Kenneth Champion; Michael D Williams
Journal:  Obes Surg       Date:  2003-12       Impact factor: 4.129

  3 in total
  10 in total

1.  Characterization of device-related interruptions in minimally invasive surgery: need for intraoperative data and effective mitigation strategies.

Authors:  James J Jung; Arash Kashfi; Sahil Sharma; Teodor Grantcharov
Journal:  Surg Endosc       Date:  2019-01-28       Impact factor: 4.584

2.  Hilar control during laparoscopic donor nephrectomy: Practice patterns in Canada.

Authors:  Thomas B Mcgregor; Premal Patel; Gabriel Chan; Alp Sener
Journal:  Can Urol Assoc J       Date:  2017-10       Impact factor: 1.862

3.  The influence of the different forms of appendix base closure on patient outcome in laparoscopic appendectomy: a randomized trial.

Authors:  Samir Delibegović; Zlatan Mehmedovic
Journal:  Surg Endosc       Date:  2017-11-02       Impact factor: 4.584

4.  Intra-operative double-stapled colorectal or coloanal anastomotic complications of laparoscopic low anterior resection for rectal cancer: double-stapled anastomotic complication could result in persistent anastomotic leakage.

Authors:  Ji Hoon Kim; Ho Young Kim; In Kyu Lee; Seung Teak Oh; Jun Gi Kim; Yoon Suk Lee
Journal:  Surg Endosc       Date:  2014-12-18       Impact factor: 4.584

5.  Continuously sutured versus linear-stapled anastomosis in robot-assisted hybrid Ivor Lewis esophageal surgery following neoadjuvant chemoradiotherapy: a single-center cohort study.

Authors:  Fiorenzo V Angehrn; Kerstin J Neuschütz; Daniel C Steinemann; Martin Bolli; Lana Fourie; Pauline Becker; Markus von Flüe
Journal:  Surg Endosc       Date:  2022-07-19       Impact factor: 3.453

6.  Renal pedicle control in laparoscopic donor nephrectomy: Evaluation of a single-center experience.

Authors:  Moayid Fallatah; Abdullah W Aldughiman; Abdulrahman S Binjawhar; Bader A Melaibary; Hossam S El-Tholoth; Abdullah S Al-Gadheeb; Ahmad Y Alzahrani; Tarek M Zahrani; Hamad S Alakrash
Journal:  Urol Ann       Date:  2022-02-15

Review 7.  Staplers or clips?: A systematic review and meta-analysis of vessel controlling devices for renal pedicle ligation in laparoscopic live donor nephrectomy.

Authors:  Yu Liu; Zhongli Huang; Yuntian Chen; Banghua Liao; Deyi Luo; Xiaoshuai Gao; Kunjie Wang; Hong Li
Journal:  Medicine (Baltimore)       Date:  2018-11       Impact factor: 1.889

8.  A comparison of renal vascular control techniques during laparoscopic nephrectomy.

Authors:  Gokhan Koc; Gokhan Rahmi Ekin; Batuhan Ergani; Yusuf Ozlem Ilbey
Journal:  J Minim Access Surg       Date:  2021 Apr-Jun       Impact factor: 1.407

9.  Stapler Malfunctions in Bariatric Surgery: An Analysis of the MAUDE Database.

Authors:  Benjamin Clapp; Alexander Schrodt; Maria Ahmad; Ellen Wicker; Nishtha Sharma; Andres Vivar; Brian Davis
Journal:  JSLS       Date:  2022 Jan-Mar       Impact factor: 1.789

10.  Characteristics of Medical Adverse Events/Near Misses Associated With Laparoscopic/Thoracoscopic Surgery: A Retrospective Study Based on the Japanese National Database of Medical Adverse Events.

Authors:  Takashige Abe; Sachiyo Murai; Yasuyuki Nasuhara; Nobuo Shinohara
Journal:  J Patient Saf       Date:  2019-12       Impact factor: 2.844

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.