Katarzyna Jończyk-Potoczna1, Marta Frankiewicz, Małgorzata Warzywoda, Krzysztof Strzyżewski, Bogdan Pawlak.
Abstract
BACKGROUND: A suspicion of ventriculo-peritoneal shunt failure is classified as the most common indication for CT in children with hydrocephalus. The main target of the study was to evaluate the diagnostic value of a low-dose protocol and to compare a total DLP received by patients in compared protocols. MATERIAL/Entities:
Keywords: CT; children; hydrocephalus; low-dose protocol
Year: 2012 PMID: 22802860 PMCID: PMC3389960 DOI: 10.12659/pjr.882575
Source DB: PubMed Journal: Pol J Radiol ISSN: 1733-134X
Specifications of protocols.
| CareDose 4D function – on | CareDose 4D function – on |
| CareDose 4D function – off | CareDose 4D function – off |
Figure 1.Nonenhanced computed tomography. (A) axial brain CT, position of the ventriculo – peritoneal shunt catheter, (B) course of the catheters outside the skull in VRT. Both images were obtained in the low-dose protocol.
Comparison of the total doses of DLP* received by patients in both groups covered by the analysis.
| Max. | 1333 | 2087 | 754 (36%) |
| Min. | 156 | 234 | 78 (33%) |
| Mean | 563 | 933 | 370 (40%) |
DLP is a measure of accumulated dose (total energy of radiation) received by the patient. The unit of this index is mGy × cm. It incorporates not only the mean radiation dose in the visualised volume (CTDIvol) but also the length of the scanned area L.
Comparison of total doses of DLP received by patients during head CTs with standard and low-dose protocols.
| A | 687 | 1193 | 506 | 1 | 2 |
| B | 844 | 1333 | 489 | 6 | 1 |
| C | 601 | 1089 | 488 | 2 | 2 |
| D | 247 | 311 | 64 | 3 | 1 |
| E | 225 | 837 | 612 | 3 | 3 |
| F | 208 | 380 | 172 | 1 | 1 |
| G | 680 | 1284 | 604 | 1 | 1 |
| H | 439 | 593 | 154 | 1 | 1 |
| I | 208 | 397 | 189 | 2 | 1 |
The difference in doses between the low – dose protocol and the standard protocol in the same patient.
| A | 687 | 1193 | 506 | 42% |
| B | 844 | 1333 | 489 | 37% |
| C | 601 | 1089 | 488 | 45% |
| D | 247 | 311 | 64 | 21% |
| E | 225 | 837 | 612 | 73% |
| F | 208 | 380 | 172 | 45% |
| G | 680 | 1284 | 604 | 47% |
| H | 439 | 593 | 154 | 26% |
| I | 208 | 397 | 189 | 48% |
Figure 2.Nonenhanced computed tomography, axial and sagittal brain CT. (A) image in the low-dose protocol, (B) image in the standard protocol.