| Literature DB >> 2279489 |
E Busemann Sokole1, T D Cradduck, J J Erickson.
Abstract
Ten gated cardiac software phantoms, representing normal and abnormal clinical conditions, were transferred to 9 different computer systems and tested with 11 cardiac programmes. Problems of inappropriate data format were encountered when analysing the phantoms on other systems. The global left ventricular ejection fraction (LVEF) values resulting from the different programmes were compared. Significant programme differences were found, the programmes falling generally into two groups. Few LVEF outlier values were identified. Full functional assessment of cardiac phantoms requires a set of different views which together form the phantom. Application of software phantoms for programme assessment and training requires prudence and attention to the acquired data format, frame time, and gating method for the appropriate use of phantoms.Entities:
Mesh:
Year: 1990 PMID: 2279489 DOI: 10.1007/bf00811435
Source DB: PubMed Journal: Eur J Nucl Med ISSN: 0340-6997