Literature DB >> 22784913

Acute posttraumatic catastrophic failure of a second-generation, highly cross-linked ultra-high-molecular-weight polyethylene patellar component.

Matthew J Goldstein1, Michael P Ast, Frank R Dimaio.   

Abstract

A 74-year-old man underwent total knee arthroplasty in 1996 and subsequent revision total knee arthroplasty for aseptic loosening in 2005. During revision, an all-polyethylene, 7-mm, second-generation, highly cross-linked ultra-high-molecular-weight polyethylene patellar component (Durasul; Zimmer, Warsaw, Indiana) was used. The patient recovered well, but he presented with severe acute pain after a popping feeling was detected during a game of golf in postoperative year 4. Radiographs demonstrated a fracture of the patellar component. At re-revision surgery, polyethylene fragments were encountered without visual evidence of wear. Appropriate rotational and axial alignment of the components was confirmed. Patellar revision was performed. To the authors' knowledge, this is the first report of an acute posttraumatic catastrophic fracture of a second-generation, highly cross-linked ultra-high-molecular-weight polyethylene patellar component. Although no malposition of the components was noted, abnormal tensile forces across the patellar component can be transmitted by altered patellofemoral kinematics. This combination could lead to a small surface crack that propagates deeper into the component with continued impaction and an eventual acute fracture of the entire component. The patellar component failed in a piecemeal fashion following acute trauma. Although many advantages exist to the enhancement of polyethylene patellar components, including the potential for improved wear characteristics, cross-linking has reduced resistance to crack initiation and propagation. Therefore, fracture risk should be considered when using second-generation, highly cross-linked ultra-high-molecular-weight polyethylene. Copyright 2012, SLACK Incorporated.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22784913     DOI: 10.3928/01477447-20120621-35

Source DB:  PubMed          Journal:  Orthopedics        ISSN: 0147-7447            Impact factor:   1.390


  5 in total

1.  Patellofemoral kinematics during deep knee flexion after total knee replacement: a computational simulation.

Authors:  Chang-Hung Huang; Lin-I Hsu; Kun-Jhih Lin; Ting-Kuo Chang; Cheng-Kung Cheng; Yung-Chang Lu; Chen-Sheng Chen; Chun-Hsiung Huang
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-01-03       Impact factor: 4.342

Review 2.  Safe clinical use of carbon nanotubes as innovative biomaterials.

Authors:  Naoto Saito; Hisao Haniu; Yuki Usui; Kaoru Aoki; Kazuo Hara; Seiji Takanashi; Masayuki Shimizu; Nobuyo Narita; Masanori Okamoto; Shinsuke Kobayashi; Hiroki Nomura; Hiroyuki Kato; Naoyuki Nishimura; Seiichi Taruta; Morinobu Endo
Journal:  Chem Rev       Date:  2014-04-10       Impact factor: 60.622

Review 3.  Highly cross-linked polyethylene may not have an advantage in total knee arthroplasty.

Authors:  Vasileios I Sakellariou; Peter Sculco; Lazaros Poultsides; Timothy Wright; Thomas P Sculco
Journal:  HSS J       Date:  2013-08-10

4.  Fracture of Two Moderately Cross-Linked Polyethylene Tibial Inserts in a TKR Patient.

Authors:  Matthew G Teeter; James P McAuley; Douglas D Naudie
Journal:  Case Rep Orthop       Date:  2014-01-08

Review 5.  Risk Factors for Knee Injury in Golf: A Systematic Review.

Authors:  Matthew L Baker; Devakar R Epari; Silvio Lorenzetti; Mark Sayers; Urs Boutellier; William R Taylor
Journal:  Sports Med       Date:  2017-12       Impact factor: 11.136