Literature DB >> 22780872

Effect of dental implant diameter on fatigue performance. Part II: failure analysis.

Keren Shemtov-Yona1, Daniel Rittel, Eli E Machtei, Liran Levin.   

Abstract

PURPOSE: The purpose of this study was to perform fracture mode analysis for in vitro failed implants in order to evaluate the relation between the fracture mode obtained and the implants' fatigue behavior.
MATERIALS AND METHODS: Eighty fractured dental implants were analyzed after being tested for fatigue performance. A macroscopic failure analysis was performed, which evaluated and located the fracture modes obtained, followed by a microscopic failure analysis comprising a detailed scanning electron microscopy (SEM) fractographic analysis.
RESULTS: Four distinctive fracture loci were identified and macrofracture mode analysis was performed, showing that all 5-mm implants that fractured were fractured at the abutment neck and screw. In the 3.75-mm group, 44.4% were fractured at the implant neck and 55.5% at the implants second thread. Fifty-two percent of the 3.3-mm fractured implants had it at the implants second tread and 48% at the implants third thread. The implant's metallographic sections revealed that the different fracture loci were located where thin metal cross sections and sharp notches coexist. Using SEM, we were able to characterize the failure micromechanisms and fatigue characterization as transgranular fracture and arrays of secondary parallel microcracks at relatively low magnifications and classic fatigue striations at much higher magnifications.
CONCLUSIONS: The results of this study indicate that proper implant design is crucial to ensure long-term fatigue performance for dental implants. The combination of sharp notches (thread) and narrow metal cross section is quite deleterious for fatigue resistance.
© 2012 Wiley Periodicals, Inc.

Entities:  

Keywords:  implant design; scanning electron microscopy; stress concentration; striation

Mesh:

Substances:

Year:  2012        PMID: 22780872     DOI: 10.1111/j.1708-8208.2012.00476.x

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  7 in total

1.  [Clinical classification and treatment decision of implant fracture].

Authors:  Y Li; H J Yu; L X Qiu
Journal:  Beijing Da Xue Xue Bao Yi Xue Ban       Date:  2022-02-18

2.  Fatigue of Dental Implants: Facts and Fallacies.

Authors:  Keren Shemtov-Yona; Daniel Rittel
Journal:  Dent J (Basel)       Date:  2016-05-24

3.  SEM Analysis and Management of Fracture Dental Implant.

Authors:  Archana Singh; Ankita Singh; Rajul Vivek; T P Chaturvedi; Pankaj Chauhan; Shruti Gupta
Journal:  Case Rep Dent       Date:  2013-04-23

4.  Spontaneous fracture of cranioplastic titanium implants without head trauma in an adult: A case report.

Authors:  Ying Jiang; Yun-Kun Wang; Ming-Kun Yu
Journal:  Int J Surg Case Rep       Date:  2016-04-28

Review 5.  An Overview of the Mechanical Integrity of Dental Implants.

Authors:  Keren Shemtov-Yona; Daniel Rittel
Journal:  Biomed Res Int       Date:  2015-10-25       Impact factor: 3.411

6.  Long-term cumulative survival and mechanical complications of single-tooth Ankylos Implants: focus on the abutment neck fractures.

Authors:  Hye Won Shim; Byoung-Eun Yang
Journal:  J Adv Prosthodont       Date:  2015-12-30       Impact factor: 1.904

Review 7.  Implant fractures: Rare but not exceptional.

Authors:  Sahitya Sanivarapu; Srinivas Moogla; Rupa Sruthi Kuntcham; Lakshmi Kanth Kolaparthy
Journal:  J Indian Soc Periodontol       Date:  2016 Jan-Feb
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.