Ultrasound confirmation of endotracheal tube position in neonates.

Debra Dennington1, Payam Vali, Neil N Finer, Jae H Kim.   

Abstract

BACKGROUND: The placement of the endotracheal tube (ETT) in neonates is a challenging procedure that currently requires timely confirmation of tip placement by radiographic imaging.
OBJECTIVE: We sought to determine if bedside ultrasound (US) could demonstrate ETT tip location in preterm and term newborns and offer a quick alternative method of ETT positioning.
METHODS: We conducted a prospective pilot study of 30 newborns admitted to the UC San Diego Medical Center who had their ETT placement confirmed by chest radiographs. After a radiograph, each infant had a US exam with a 13-MHz linear transducer on a portable US machine. To assist localization, gentle longitudinal movement of the ETT of less than 0.5 cm was performed. Measurements from the tip of the ETT tip to the carina were made on chest radiograph and midsagittal US images.
RESULTS: Study infants had a mean gestational age of 30.2 ± 4.9 (SD) weeks and mean birth weight of 1,595.2 ± 862 g. US images were taken a mean 2.9 ± 2.2 h after radiographs. Data from 2 infants were excluded for poor radiograph image quality and extreme outlier values. The ETT was visualized by US in all newborns examined. We observed a good correlation between ETT tip-to-carina distance on US and radiograph (r(2) = 0.68) with minimal bias. Each study took less than 5 min to obtain without any clinical deterioration.
CONCLUSIONS: Bedside US can visualize the anatomic position of the ETT position in preterm and term infants but further validation is required before routine clinical implementation.
Copyright © 2012 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2012        PMID: 22777009     DOI: 10.1159/000338585

Source DB:  PubMed          Journal:  Neonatology        ISSN: 1661-7800            Impact factor:   4.035


  9 in total

Review 1.  Ultrasonography for endotracheal tube position in infants and children.

Authors:  Pooja Jaeel; Mansi Sheth; Jimmy Nguyen
Journal:  Eur J Pediatr       Date:  2017-01-13       Impact factor: 3.183

2.  The concordance of ultrasound technique versus X-ray to confirm endotracheal tube position in neonates.

Authors:  R Chowdhry; B Dangman; J M B Pinheiro
Journal:  J Perinatol       Date:  2015-01-22       Impact factor: 2.521

Review 3.  Sonographic Evaluation of the Endotracheal Tube Position in the Neonatal Population: A Comprehensive Review and Meta-Analysis.

Authors:  Sabrina Congedi; Federica Savio; Maria Auciello; Sabrina Salvadori; Daniel Nardo; Luca Bonadies
Journal:  Front Pediatr       Date:  2022-06-02       Impact factor: 3.569

Review 4.  Neonatal respiratory distress syndrome: Chest X-ray or lung ultrasound? A systematic review.

Authors:  Matthew Hiles; Anne-Marie Culpan; Catriona Watts; Theresa Munyombwe; Stephen Wolstenhulme
Journal:  Ultrasound       Date:  2017-01-29

Review 5.  WITHDRAWN: Techniques to ascertain correct endotracheal tube placement in neonates.

Authors:  Georg M Schmölzer; Charles Christoph C Roehr
Journal:  Cochrane Database Syst Rev       Date:  2018-07-05

6.  A feasibility study on bedside upper airway ultrasonography compared to waveform capnography for verifying endotracheal tube location after intubation.

Authors:  Osman Adi; Tan Wan Chuan; Manikam Rishya
Journal:  Crit Ultrasound J       Date:  2013-07-04

7.  Optimal Line and Tube Placement in Very Preterm Neonates: An Audit of Practice.

Authors:  Daragh Finn; Hannah Kinoshita; Vicki Livingstone; Eugene M Dempsey
Journal:  Children (Basel)       Date:  2017-11-17

8.  Comparative Evaluation of Difficult Intubation Predictors in Children Under Two Years of Ages.

Authors:  Sedighe Shahhosseini; Mohammad Montasery; Mohammadali Saadati; Amir Shafa
Journal:  Anesth Pain Med       Date:  2021-12-08

9.  Ultrasound-Guided Selective Bronchial Intubation: A Feasibility Study in Pediatric Animal Model.

Authors:  Sara Hora Gomes; Alice Miranda; José Miguel Pêgo; Patrício S Costa; Jorge Correia-Pinto
Journal:  Front Med (Lausanne)       Date:  2022-06-15
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.