Literature DB >> 22773019

Laser targeting with C-arm fluoroscopy: effect on image acquisition and radiation exposure.

Franklin D Shuler1, Justin L Daigre, Danh Pham, Vincent L Kish.   

Abstract

OBJECTIVES: Reducing unnecessary radiation exposure from medical imaging is paramount. This study assessed whether a laser aiming guide for C-arm fluoroscopy reduced the number of exposures needed to obtain an acceptable image, thereby reducing total fluoroscopy time for hip, knee, and ankle fluoroscopy.
METHODS: An obese cadaver was placed supine on a radiolucent surgical table. Images were obtained by licensed radiologic technologists using a calibrated OEC 9900 Elite C-arm with laser targeting (LT) and without LT (NLT). Dosimeters were placed 1, 3, and 6 ft (30.5, 91.5, and 183 cm) away from the center of the C-arm at 90-degree angles at 2 levels, simulating thyroid and gonadal exposure. Posterior-anterior (PA) images of the bilateral lower extremities were obtained with each technician acquiring 24 centered images (hip, knee, and ankle) using both LT and NLT C-arm fluoroscopy.
RESULTS: Total fluoroscopy time was reduced by 19% when using LT with a 39% reduction for the knee and a 29% reduction for the ankle. The addition of LT improved the likelihood of obtaining a centered image for knees and ankles but not for hips. The gonadal dosimetry data were significantly higher than the thyroid dosimetry badges at 1 ft. At the 3-ft zone, only trace amounts of radiation were detected; the 6-ft zone reported no radiation exposure in either group.
CONCLUSIONS: LT helped with imaging knees and ankles with statistically significant reductions in fluoroscopy time and a statistically significant improvement of image quality defined as obtaining a centered PA image faster. The dosimetry badges detected minimal exposure at 3 ft and no detectable exposure at 6 ft at both levels.

Entities:  

Mesh:

Year:  2013        PMID: 22773019     DOI: 10.1097/BOT.0b013e31826625df

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  3 in total

Review 1.  Intraoperative radiation safety in orthopaedics: a review of the ALARA (As low as reasonably achievable) principle.

Authors:  Daniel J Kaplan; Jay N Patel; Frank A Liporace; Richard S Yoon
Journal:  Patient Saf Surg       Date:  2016-12-12

2.  The Effect of C-Arm Mobility and Field of Vision on Radiation Exposure in the Treatment of Proximal Femoral Fractures: A Randomized Clinical Trial.

Authors:  Mahmut Kalem; Kerem Başarır; Hakan Kocaoğlu; Ercan Şahin; Hakan Kınık
Journal:  Biomed Res Int       Date:  2018-03-27       Impact factor: 3.411

Review 3.  Radiation safety for pain physicians: principles and recommendations.

Authors:  Sewon Park; Minjung Kim; Jae Hun Kim
Journal:  Korean J Pain       Date:  2022-04-01
  3 in total