| Literature DB >> 22772024 |
S Shcherbinin1, S Chamoiseau, A Celler.
Abstract
We investigated the quantitative accuracy of the model-based dual-isotope single-photon emission computed tomography (DI-SPECT) reconstructions that use Klein-Nishina expressions to estimate the scattered photon contributions to the projection data. Our objective was to examine the ability of the method to recover the absolute activities pertaining to both radiotracers: Tc-99m and I-123. We validated our method through a series of phantom experiments performed using a clinical hybrid SPECT/CT camera (Infinia Hawkeye, GE Healthcare). Different activity ratios and different attenuating media were used in these experiments to create cross-talk effects of varying severity, which can occur in clinical studies. Accurate model-based corrections for scatter and cross-talk with CT attenuation maps allowed for the recovery of the absolute activities from DI-SPECT/CT scans with errors that ranged 0-10% for both radiotracers. The unfavorable activity ratios increased the computational burden but practically did not affect the resulting accuracy. The visual analysis of parathyroid patient data demonstrated that our model-based processing improved adenoma/background contrast and enhanced localization of small or faint adenomas.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22772024 DOI: 10.1088/0031-9155/57/15/4755
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609