Monte Carlo simulation and measurement of radiation leakage from applicators used in external electron radiotherapy.

Tomohiro Shimozato1, Kuniyasu Okudaira, Hiraku Fuse, Katsuyoshi Tabushi.   

Abstract

External electron radiotherapy is performed using a cone or applicator to collimate the beam. However, because of a trade-off between collimation and scattering/bremsstrahlung X-ray production, applicators generate a small amount of secondary radiation (leakage). We investigate the peripheral dose outside the radiation field of a Varian-type applicator. The dose and fluence outside the radiation field were analyzed in a detailed Monte Carlo simulation. The differences between the calculation results and data measured in a water phantom in an ionization chamber were less than ±1% in regions more than 3 mm below the surface of the phantom and at the depth of dose maximum. The calculated fluence was analyzed inside and outside the radiation field on a plane just above the water phantom surface. Changing the electron energy affected the off-axis fluence distribution outside the radiation field; however, the size of the applicator had little effect on this distribution. For each energy, the distributions outside the radiation field were similar to the dose distribution at shallow depths in the water phantom. The effect of secondary electrons generation by photon transmission through the alloy making up the lowest scraper was largest in the region from the field edge to directly below the cutout and at higher beam energies. The results of the Monte Carlo simulation confirm that the peripheral dose outside the field is significantly affected by radiation scattered or transmitted from the applicator, and the effect increases with the electron energy.
Copyright © 2012 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22771332     DOI: 10.1016/j.ejmp.2012.06.006

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  5 in total

1.  Excessive applicator radiation leakage for a common therapeutic kilovoltage system.

Authors:  Brad Beeksma; Joerg Lehmann
Journal:  Br J Radiol       Date:  2018-11-15       Impact factor: 3.039

2.  Experimental assessment of out-of-field dose components in high energy electron beams used in external beam radiotherapy.

Authors:  Mohamad M Alabdoaburas; Jean-Pierre Mege; Jean Chavaudra; Jérémi Vũ Bezin; Atilla Veres; Florent de Vathaire; Dimitri Lefkopoulos; Ibrahima Diallo
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

3.  Nontarget and Out-of-Field Doses from Electron Beam Radiotherapy.

Authors:  Natalia Matuszak; Marta Kruszyna-Mochalska; Agnieszka Skrobala; Adam Ryczkowski; Piotr Romanski; Igor Piotrowski; Katarzyna Kulcenty; Wiktoria Maria Suchorska; Julian Malicki
Journal:  Life (Basel)       Date:  2022-06-08

4.  Out-of-field doses and neutron dose equivalents for electron beams from modern Varian and Elekta linear accelerators.

Authors:  Carlos E Cardenas; Paige L Nitsch; Rajat J Kudchadker; Rebecca M Howell; Stephen F Kry
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

5.  Radiation leakage dose from Elekta electron collimation system.

Authors:  Garrett M Pitcher; Kenneth R Hogstrom; Robert L Carver
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.