| Literature DB >> 22770854 |
Sean T O'Brien1, Colin D Burnell, David R Hedden, Jan-M Brandt.
Abstract
A 34-year-old female patient received a cobalt-chromium (CoCr) alloy femoral head on cross-linked polyethylene total hip replacement for the revision of her fractured ceramic-on-ceramic total hip replacement. The CoCr alloy femoral head became severely worn due to third-body abrasive wear by ceramic particles that could not be removed by synovectomy or irrigation at revision surgery. Ceramic particles were found embedded in the cross-linked polyethylene liner. The CoCr alloy femoral head exhibited a total mass loss of 14.2 g and the generated wear particles triggered metallosis in the patient. The present case study suggests not revising a fractured ceramic-on-ceramic total hip replacement with a CoCr alloy femoral head and a cross-linked polyethylene liner to avoid metallosis due to third-body abrasive wear. CrownEntities:
Mesh:
Substances:
Year: 2012 PMID: 22770854 DOI: 10.1016/j.arth.2012.05.014
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757