Literature DB >> 22763907

Minute-to-minute urine flow rate variability: a new renal physiology variable.

Yoram Klein1, Mor Grinstein, Stephen M Cohn, Jacob Silverman, Moti Klein, Hanoch Kashtan, Micha Y Shamir.   

Abstract

BACKGROUND: Urine output is a surrogate for tissue perfusion and is typically measured at 1-hour intervals. Because small urine volumes are difficult to measure in urine collection bags, considerable over- or underestimation is common. To overcome these shortcomings, digital urine meters were developed. Because these monitors measure urine volume in 1-minute intervals, they provide minute-to-minute measurements of the urine flow rate (UFR). In a previous study, we observed that the minute-to-minute variability in the UFR disappeared during hypovolemia. The aim of this study was to describe the minute-to-minute variability in the UFR as a new physiological variable and to show its relationship to blood volume depletion.
METHODS: Seven adult pigs were used in this study. The UFR, minute-to-minute UFR, mean arterial blood pressure, heart rate, and base excess were measured at euvolemia and during gradual hemorrhaging (10%, 20%, and 30% of estimated blood volume). Variance and wavelet spectral analysis were used to measure the disappearance of the minute-to-minute UFR variability.
RESULTS: The UFR decreased from 2.2 ± 0.2 to 1.0 ± 0.1 mL/min after a 10% estimated blood volume loss (±1 SE, n = 7, P = 0.0348). The variance in the minute-to-minute UFR decreased from 1.4 ± 0.3 to 0.4 ± 0.1 mL/min (±1 SE, n = 7, P = 0.046).
CONCLUSIONS: The UFR and its minute-to-minute variability decrease during hemorrhaging. The variability in the UFR may be useful as an aid for the diagnosis of hypovolemia.

Entities:  

Mesh:

Year:  2012        PMID: 22763907     DOI: 10.1213/ANE.0b013e3182625813

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  3 in total

1.  Urine flow rate monitoring in hypovolemic multiple trauma patients.

Authors:  Evgeni Brotfain; Yoram Klein; Ronen Toledano; Leonid Koyfman; Dmitry Frank; Micha Y Shamir; Moti Klein
Journal:  World J Emerg Surg       Date:  2017-08-18       Impact factor: 5.469

2.  On the minute by minute variations of urine output: a study in a porcine model.

Authors:  Abraham Otero; Pablo Cardinal-Fernández; Yeny Rojas; Nicolás Nin; Leticia Martínez-Caro; Andrés Esteban; José A Lorente
Journal:  J Nephrol       Date:  2014-01-15       Impact factor: 3.902

3.  A new device to automate the monitoring of critical patients' urine output.

Authors:  Abraham Otero; Andrey Apalkov; Roemi Fernández; Manuel Armada
Journal:  Biomed Res Int       Date:  2014-01-28       Impact factor: 3.411

  3 in total