Noncontact ultrasound imaging applied to cortical bone phantoms.

J B Bulman1, K S Ganezer, P W Halcrow, Ian Neeson.   

Abstract

PURPOSE: The purpose of this paper was to take the first steps toward applying noncontact ultrasound (NCU) to the tasks of monitoring osteoporosis and quantitative ultrasound imaging (QUS) of cortical bone. The authors also focused on the advantages of NCU, such as its lack of reliance on a technologist to apply transducers and a layer of acoustical coupling gel, the ability of the transducers to operate autonomously as specified by preprogrammed software, and the likely reduction in statistical and systematic errors associated with the variability in the pressure applied by the clinician to the transmitting transducer that NCU might provide. The authors also undertook this study in order to find additional applications of NCU beyond its past limited usage in assessing the severity of third degree burns.
METHODS: A noncontact ultrasound imaging system using a pair of specially designed broadband, 1.5 MHz noncontact piezoelectric transducers and cortical bone phantoms, were used to determine bone mineral density (BMD), speed of sound (SOS), integrated response (IR), and ultrasonic transmittance. Air gaps of greater than 3 cm, two transmission and two reflection paths, and a digital signal processor were also used in the collection of data from phantoms of nominal mass densities that varied from 1.17 to 2.25 g/cm(3) and in bone mineral density from 0 to 1.7 g/cm(3).
RESULTS: Good correlations between known BMD and measured SOS, IR, and transmittance were obtained for all 17 phantoms, and methods for quantifying and minimizing sources of systematic errors were outlined. The BMD of the phantom sets extended through most of the in vivo range found in cortical bone. A total of 16-20 repeated measurements of the SOS, thickness, and IR for the phantom set that were conducted over a period of several months showed a small variation in the range of measurements of ±1%-2%. These NCU data were shown to be in agreement with similar results using contact ultrasound to be within 1%-2%. Transmittance images of cortical bone phantoms showed differences in the nominal overall BMD values of the phantoms that were large enough to be distinguished by a visual examination. A list of possible sources of errors in quantitative NCU was also included in this study.
CONCLUSIONS: The results of this paper suggest that NCU might find additional applications in medical imaging, beyond its original and only previous usage in assessing third degree burns. The fact that the authors' phantom measurements using conventional, gel coupled ultrasound are in agreement with those obtained with NCU demonstrates that in spite of large additional levels of attenuation of up to 150 dB and new error sources, NCU could have comparable levels of accuracy to those of conventional quantitative ultrasound, while providing the medical and patient comfort-related advantages of not involving direct contact.
© 2012 American Association of Physicists in Medicine.

Entities:  

Mesh:

Year:  2012        PMID: 22755697      PMCID: PMC3365917          DOI: 10.1118/1.4709598

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  38 in total

1.  An outbreak of pyodermas among neonates caused by ultrasound gel contaminated with methicillin-susceptible Staphylococcus aureus.

Authors:  K Weist; C Wendt; L R Petersen; H Versmold; H Rüden
Journal:  Infect Control Hosp Epidemiol       Date:  2000-12       Impact factor: 3.254

2.  A contact method for the assessment of ultrasonic velocity and broadband attenuation in cortical and cancellous bone.

Authors:  C M Langton; A V Ali; C M Riggs; G P Evans; W Bonfield
Journal:  Clin Phys Physiol Meas       Date:  1990-08

Review 3.  Methods of bone mineral measurement.

Authors:  P Tothill
Journal:  Phys Med Biol       Date:  1989-05       Impact factor: 3.609

4.  Use of multiple acoustic wave modes for assessment of long bones: model study.

Authors:  Alexey Tatarinov; Noune Sarvazyan; Armen Sarvazyan
Journal:  Ultrasonics       Date:  2005-03-31       Impact factor: 2.890

5.  The interaction of ultrasound with cancellous bone.

Authors:  M L McKelvie; S B Palmer
Journal:  Phys Med Biol       Date:  1991-10       Impact factor: 3.609

6.  Connectivity in human cancellous bone by three-dimensional magnetic resonance imaging.

Authors:  M Wessels; R P Mason; P P Antich; J E Zerwekh; C Y Pak
Journal:  Med Phys       Date:  1997-09       Impact factor: 4.071

7.  Microstructural insight into pedestrian pelvic fracture as assessed by high-resolution computed tomography.

Authors:  J G Snedeker; F H Walz; M H Muser; G Schroeder; T L Mueller; R Müller
Journal:  J Biomech       Date:  2005-10-25       Impact factor: 2.712

8.  Nosocomial outbreak of Klebsiella pneumoniae producing SHV-5 extended-spectrum beta-lactamase, originating from a contaminated ultrasonography coupling gel.

Authors:  O Gaillot; C Maruéjouls; E Abachin; F Lecuru; G Arlet; M Simonet; P Berche
Journal:  J Clin Microbiol       Date:  1998-05       Impact factor: 5.948

9.  The measurement of broadband ultrasonic attenuation in cancellous bone.

Authors:  C M Langton; S B Palmer; R W Porter
Journal:  Eng Med       Date:  1984-04

10.  Palangeal quantitative ultrasound, phalangeal morphometric variables, and vertebral fracture discrimination.

Authors:  G Guglielmi; C F Njeh; F de Terlizzi; D A De Serio; A Scillitani; M Cammisa; B Fan; Y Lu; H K Genant
Journal:  Calcif Tissue Int       Date:  2003-02-10       Impact factor: 4.333

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.