Literature DB >> 22754260

Simultaneous Quantitative Analysis of Olmesartan Medoxomil and Amlodipine Besylate in Plasma by High-performance Liquid Chromatography Technique.

S Shah1, A Asnani, D Kawade, S Dangre, S Arora, S Yende.   

Abstract

A rapid, simple and sensitive high-performance liquid chromatography (HPLC) method has been developed for quantification of olmesartan medoxomil (OLM) and amlodipine besylate (AM) in plasma. The assay enables the measurement of OLM and AM for therapeutic drug monitoring with a minimum detectable limit of 2 ng mL. The method involves a simple, one-step extraction procedure and analytical recovery was above 50%. The separation was performed on an analytical 250 × 4.6 mm Eurospher 100(-5) C18 column. The wavelength was set at 239 nm. The mobile phase was a mixture of acetonitrile:0.05 M ammonium acetate buffer: 0.1 mL triethylamine at pH 6.8 was selected at a flow rate of 1.0 mL min. The calibration curve for the determination of OLM and AM in plasma was linear over the range 2-2500 and 8-10,000 ng mL AM and OLM. The coefficients of variation for interday and intraday assay were found to be <15%. The method can be applied to a pharmacokinetic and pharmacodynamic study of OLM and AM in a combined dosage form.

Entities:  

Year:  2012        PMID: 22754260      PMCID: PMC3385223          DOI: 10.4103/0975-1483.96622

Source DB:  PubMed          Journal:  J Young Pharm        ISSN: 0975-1483


INTRODUCTION

Amlodipine, R, S-2-[(2-aminoethoxy)methyl]-4-(2-chlorophenyl)-ethoxycarbonyl-5-methoxy-carbonyl-6-methyl-1, 4-dihydropyridine [Figure 1], is a potent calcium channel blocker used in treatment of hypertension and angina pectoris.[1] Amlodipine is well absorbed following oral administration with peak blood concentrations occurring after 6–12 h. The bioavailability is about 60–65%. It has a prolonged terminal elimination half-life of 35–50 h, and steady-state plasma concentrations are not achieved until after 7–8 days of administration. Amlodipine is extensively metabolized in the liver; metabolites are mostly excreted in urine together with <10% of a dose as unchanged drug.[23] Several analytical methods for quantifying of amlodipine in biological fluids have been reported; such as capillary gas chromatography (GC) with electron capture detection,[45] GC with electron-impact mass spectrometry (EI-MS),[67] and high-performance liquid chromatography (HPLC).[8-12] Several HPLC procedures have been also reported for the analyses of amlodipine based on MS–MS detection by using tandem mass spectrometry.[13-16] Olmesartan medoxomil (OLM), (5-methyl-2-oxo-1,3-dioxolen-4-yl) methoxy-4-(1-hydroxy-1-methylethyl)-2-propyl-1-{4-[2-(tetrazol-5-yl)-phenyl]phenyl}methyl imidazol-5-carboxylate) [Figure 2], is a potent and selective angiotensin AT1 receptor blocker[17] which has been approved for the treatment of hypertension in the United States, Japan, and European countries. The drug contains a medoxomil ester moiety and is cleaved rapidly by an endogenous esterase to release the active metabolite olmesartan.[18] Up to date, olmesartan has been determined in plasma and other biological fluids using HPLC coupled to fluorescent detection.[1920]
Figure 1

Structure of amlodipine

Figure 2

Structure of olmesartan

Structure of amlodipine Structure of olmesartan As per literature survey, no analytical methods have been reported for the simultaneous quantitative analysis of OLM and AM in biological fluid. The aim of our study was to develop a rapid, simple, accurate, precise, sensitive, and reproducible HPLC method, which can be utilized in pharmacokinetic research.

EXPERIMENTAL

Chemicals and reagents

Olmesartan medoxomil (OLM) and amlodipine besylate (AM) were kindly supplied by Ajanta Pharma Limited (Mumbai). All chemicals and reagents were HPLC grade, methanol and acetonitrile, ammonium acetate, ethylene diamine tetra acetic acid (EDTA) (Merck, India).

Instruments and chromatographic conditions

A Shimadzu HPLC system (Japan) employed consisted of a model LC-20 AT pump, a model Rheodyne 7125 injector, and a model SPD-20A UV detector. The separation was performed on an analytical 250 × 4.6 mm Eurospher 100-5 C18 (5 μm, particle size) column. The wavelength was set at 239 nm. The mobile phase was a mixture of acetonitrile:0.05 M ammonium acetate buffer: 0.1 mL triethylamine at pH 6.8 was selected at a flow rate of 1.0 mL/min. The mobile phase was prepared daily and degassed by ultrasonification before use.

Preparation of standard stock solution

The standard working concentration of OLM (10 μg/mL) and AM (10 μg/mL) were prepared in the mobile phase.

Biological samples

The prepared suspension containing AM (0.4 mg/mL) and OLM (1.6 mg/mL) in 0.2% tween 80 were administered orally to Wistar rats (180–220 g) at a dose of 2 mg/kg (AM) and 8 mg/kg (OLM) body weight after overnight fasting. The blood samples were collected from the retrobulbar plexus into anticoagulant-treated polypropylene tubes from 0 to 12 h after drug administration. Blood samples collected were centrifuged immediately to separate the plasma. The plasma samples collected were stored at –20°C. The plasma samples then processed for drug recovery via solvent extraction.

Samples preparation (extraction procedure)

To 1000 μL of plasma samples in a borosilicate glass tube were added 5 mL of HPLC grade acetonitrile. After vortex mixing for 10 min at room temperature, the samples were centrifuged at 10,500 rpm for 10 min. The upper organic layer was transferred to a glass container and evaporated inside a vacuum oven at 40°C. The dry residue was dissolved in 1 mL of the mobile phase. The mixture was sonicated well for 10 min, and 20 μL of this solution was injected into liquid chromatography.

Stability

Stability assessments were defined as the resistance to change in concentration of an analyte in plasma under specified conditions. The stability of OLM and AM in rat plasma was investigated by adding known amount of drug to blank plasma samples to give concentrations of 10, 200, and 500 ng/mL. These were stored at -20°C and aliquots taken for 1 and 4 weeks for analysis.

Calibration curve

Linearity of instrument standard response was determined for each compound with different concentration calibration standards (final concentration ranging from 2 to 2500 ng/mL for AM and 8 to 10,000 ng/mL for OLM, with drug ratio maintaining constant 1:4). The calibration curves were constructed by plotting individual analyte peak area ratio as compared to the corresponding concentration and fitting these data in regression analysis. The calibration curves were constructed by the weighted regression method (1/x) of the peak area of OLM or AM vs. actual concentrations.

RESULTS AND DISCUSSION

Method development

Considering the efficiency of HPLC, an attempt has been made to develop a simple, accurate, precise, rapid, and economic method for the simultaneous estimation of OLM and AM in the biological method. Thus, the method described enables to the quantification of OLM and AM. The advantages lie in the simplicity of sample preparation and the low costs of reagents used. For the HPLC method development mobile phase consists of acetonitrile:0.05 M ammonium acetate buffer:0.1 mL triethylamine at pH 6.8 at a flow rate of 1.0 mL/min. The optimum wavelength selected was 239 nm. Under these chromatographic conditions described, OLM and AM peaks were well resolved. Endogenous plasma components did not give any interfering peaks. The average retention times of OLM and AM were 3.1 and 5.0 min, respectively. Chromatogram obtained after drug administration of OLM and AM in Wistar rats for blank plasma and plasma samples collected after 2.5, 6 and 10 h are shown in Figure 3.
Figure 3

Chromatogram of blank plasma, OLM, and AM after oral administration in rat plasma

Chromatogram of blank plasma, OLM, and AM after oral administration in rat plasma The calibration curve for the determination of OLM and AM in plasma was linear over the range 2–2500 and 8–10,000 ng/mL AM and OLM. The linearity of this method was statistically confirmed. For each calibration curve, the intercept was not statistically different from zero. The correlation coefficients (R) for calibration curves were equal to or better than 0.997. The relative standard deviation (RSD) values of the slope were equal to or better than 5%. For each point of calibration standards, the concentrations were recalculated from the equation of the linear regression curves.

Recovery study

The relative analytical recovery for plasma at three different concentrations of OLM and AM was determined. Known amounts of drug were added to drug-free plasma in concentrations ranging from 0.002 to 10 μg/mL. The average recovery was in between 50% and 90% for a linear concentration range. The results are given in Tables 1 and 2 for the intraday and interday study.
Table 1

Recovery studies for olmesartan medoxomil and amlodipine besylate (intraday) (RP-HPLC)

Table 2

Recovery studies for OLM and AM (interday) (RP-HPLC)

Recovery studies for olmesartan medoxomil and amlodipine besylate (intraday) (RP-HPLC) Recovery studies for OLM and AM (interday) (RP-HPLC)

Stability study

In the short-term stability studies, precision and accuracy for the analyte was <15% and <8% for OLM, <10% and 8% for AM, respectively. In the long-term stability studies, precision and accuracy for the analyte was <7% and <15% for OLM, <12% and 18% for AM, respectively. The stability tests indicated that OLM and AM are stable in rat plasma over the period of 1 month. In addition, the stock solution remains stable for a duration of 1 month when stored at 4°C expressed by the % variation <2% for both OLM and AM. The results are given in Tables 3 and 4 for OLM and AM, respectively.
Table 3

Stability of OLM in rat plasma (RP-HPLC)

Table 4

Stability of AM in rat plasma (RP-HPLC)

Stability of OLM in rat plasma (RP-HPLC) Stability of AM in rat plasma (RP-HPLC)

Limit of detection

The limit of detection (LOD) was defined, as the amlodipine concentration that produced a signal-to-noise ratio greater than 3. The LOD in plasma was 2 ng/mL for AM and 8 ng/mL based upon this criterion. At this level, the RSD was lower than 15%.

Accuracy and precision

Intraday and interday assay performed to evaluate precision (% RSD) and accuracy. The coefficient of variation for intraday studies was between 2–14% and 3–17% for interday in rat plasma. The result suggests that within run and between run experiments, the precision and accuracy for the analyte met the acceptance criteria. It does not require tedious derivatization or specialized detectors, making it readily amendable to routine laboratory use. However, this method is sensitive enough for drug monitoring and other purposes such as pharmacokinetic studies. We assessed the precision of the method by repeated analysis of plasma specimens containing known concentrations of amlodipine and olmesartan. The coefficients of variation were less than 15%, which is acceptable for the routine measurement of OLM and AM. The results for intraday and interday precision and accuracy studies are given in Tables 5 and 6.
Table 5

Data for accuracy and precision studies (Intraday) (RP-HPLC)

Table 6

Data for accuracy and precision studies (interday) (RP-HPLC)

Data for accuracy and precision studies (Intraday) (RP-HPLC) Data for accuracy and precision studies (interday) (RP-HPLC)

Pharmacokinetic study

The results of pharmacokinetics analysis of OLM and AM are mentioned in Table 7. The mean plasma concentration–time curve of OLM and AM shows: Plasma concentration reached a maximum 2 h and 6 h for OLM and AM, after dosing with a level of 8 and 2 mg kg body weight of OLM and AM, respectively. The value of pharmacokinetic parameters such as t1/2, Cmax and kel obtained reveals that there was less difference in the profile of a individual drug when administered simultaneously in a single oral dosage form. These results are agreement with the previous reports.[192122 Plasma concentration–time profile curve following a single oral administration of OLM and AM in a combined dosage form in Wistar rats is shown in Figure 4.
Table 7

Pharmacokinetic parameters of a single oral administration of OLM and AM in combined dosage form in Wistar rats

Figure 4

Plasma time profile curve

Pharmacokinetic parameters of a single oral administration of OLM and AM in combined dosage form in Wistar rats Plasma time profile curve

CONCLUSION

The HPLC method presented is direct, simple, selective, reproducible, sensitive, and linear. The procedure was successfully applied to the simultaneous determination of the studied compounds in biological fluid without any interference from the additives and endogenous substances. The procedure was fully validated to meet the requirements of the food and drug administration center for drug evaluation and research guidelines for bioanalytical method validation.[23] These methods are well-suited for routine application in the quality control laboratories and clinical laboratories because of the simplicity, economic, accuracy, sensitivity, and reproducibility. The HPLC method is applicable to pharmacokinetics studies of OLM and AM in rats. This method can also be used to study the mechanism of metabolism of OLM and AM.[24]
  22 in total

1.  Applications of liquid chromatography-tandem mass spectrometry in drug and biomedical analyses.

Authors:  D Zhong; X Chen; J Gu; X Li; J Guo
Journal:  Clin Chim Acta       Date:  2001-11       Impact factor: 3.786

2.  Enantiomeric determination of amlodipine in human plasma by liquid chromatography coupled to tandem mass spectrometry.

Authors:  B Streel; C Lainé; C Zimmer; R Sibenaler; A Ceccato
Journal:  J Biochem Biophys Methods       Date:  2002-12-31

3.  Pharmacological and pharmacokinetic study of olmesartan medoxomil in animal diabetic retinopathy models.

Authors:  Hiroaki Nakamura; Tatsuya Inoue; Naohisa Arakawa; Yuki Shimizu; Yasushi Yoshigae; Izumi Fujimori; Emi Shimakawa; Tohru Toyoshi; Tomihisa Yokoyama
Journal:  Eur J Pharmacol       Date:  2005-04-11       Impact factor: 4.432

4.  Adding hydrochlorothiazide to olmesartan dose dependently improves 24-h blood pressure and response rates in mild-to-moderate hypertension.

Authors:  Lorenz Sellin; Johannes Stegbauer; Petra Laeis; Lars C Rump
Journal:  J Hypertens       Date:  2005-11       Impact factor: 4.844

5.  Simultaneous determination and pharmacokinetic study of metformin and rosiglitazone in human plasma by HPLC-ESI-MS.

Authors:  Lingyun Chen; Zhifeng Zhou; Mei Shen; Ande Ma
Journal:  J Chromatogr Sci       Date:  2011-02       Impact factor: 1.618

6.  Coupled-column chromatography on a Chiral-AGP phase for determination of amlodipine enantiomers in human plasma: an HPLC assay with electrochemical detection.

Authors:  M Josefsson; B Norlander
Journal:  J Pharm Biomed Anal       Date:  1996-11       Impact factor: 3.935

7.  Amlodipine bioequivalence study: quantification by liquid chromatography coupled to tandem mass spectrometry.

Authors:  M Carvalho; C H Oliveira; G D Mendes; M Sucupira; M E Moraes; G De Nucci
Journal:  Biopharm Drug Dispos       Date:  2001-12       Impact factor: 1.627

8.  Gas and liquid chromatographic analyses of nimodipine calcium antagonist in blood plasma and cerebrospinal fluid.

Authors:  G J Krol; A J Noe; S C Yeh; K D Raemsch
Journal:  J Chromatogr       Date:  1984-01-13

9.  Simple and rapid HPLC method for determination of amlodipine in human serum with fluorescence detection and its use in pharmacokinetic studies.

Authors:  Gh Bahrami; Sh Mirzaeei
Journal:  J Pharm Biomed Anal       Date:  2004-09-21       Impact factor: 3.935

Review 10.  Clinical pharmacokinetics of amlodipine.

Authors:  P A Meredith; H L Elliott
Journal:  Clin Pharmacokinet       Date:  1992-01       Impact factor: 6.447

View more
  1 in total

1.  Development and Validation of a Chromatography Method Using Tandem UV/Charged Aerosol Detector for Simultaneous Determination of Amlodipine Besylate and Olmesartan Medoxomil: Application to Drug-Excipient Compatibility Study.

Authors:  Ariadne M Brondi; Jerusa S Garcia; Marcello G Trevisan
Journal:  J Anal Methods Chem       Date:  2017-12-17       Impact factor: 2.193