BACKGROUND: Transcranial Magnetic Stimulation (TMS) is based on a changing magnetic field inducing an electric field in the brain. Conventionally, the TMS coil is mounted to a static holder and the subject is asked to avoid head motion. Additionally, head resting frames have been used. In contrast, our robotized TMS system employs active motion compensation (MC) to maintain the correct coil position. OBJECTIVE/HYPOTHESIS: We study the effect of patient motion on TMS. In particular, we compare different coil positioning techniques with respect to the induced electric field.
METHODS: We recorded head motion for six subjects in three scenarios: (a) avoiding head motion, (b) using a head rest, and (c) moving the head freely. Subsequently, the motion traces were replayed using a second robot to move a sensor to measure the electric field in the target region. These head movements were combined with 2 types of coil positioning: (1) using a coil holder and (2) using robotized TMS with MC.
RESULTS: After 30 min the induced electric field was reduced by 32.0% and 19.7% for scenarios (1a) and (1b), respectively. For scenarios (2a)-(2c) it was reduced by only 4.9%, 1.4% and 2.0%, respectively, which is a significant improvement (P < 0.05). Furthermore, the orientation of the induced field changed by 5.5°, 7.6°, 0.4°, 0.2°, 0.2° for scenarios (1a)-(2c).
CONCLUSION: While none of the scenarios required rigid head fixation, using a simple holder to position a coil during TMS can lead to substantial deviations in the induced electric field. In contrast, robotic motion compensation results in clinically acceptable positioning throughout treatment.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22749687     DOI: 10.1016/j.brs.2012.06.002

Source DB:  PubMed          Journal:  Brain Stimul        ISSN: 1876-4754            Impact factor:   8.955


  8 in total

1.  Nonphysiological factors in navigated TMS studies; confounding covariates and valid intracortical estimates.

Authors:  Sein Schmidt; Rouven Bathe-Peters; Robert Fleischmann; Maria Rönnefarth; Michael Scholz; Stephan A Brandt
Journal:  Hum Brain Mapp       Date:  2014-08-29       Impact factor: 5.038

2.  Accuracy of robotic coil positioning during transcranial magnetic stimulation.

Authors:  Stefan M Goetz; I Cassie Kozyrkov; Bruce Luber; Sarah H Lisanby; David L K Murphy; Warren M Grill; Angel V Peterchev
Journal:  J Neural Eng       Date:  2019-09-17       Impact factor: 5.379

3.  Cortical representation of auricular muscles in humans: A robot-controlled TMS mapping and fMRI study.

Authors:  Jonna Meincke; Manuel Hewitt; Markus Reischl; Rüdiger Rupp; Carsten Schmidt-Samoa; David Liebetanz
Journal:  PLoS One       Date:  2018-07-27       Impact factor: 3.240

4.  A landmark-based approach to locate symptom-specific transcranial magnetic stimulation targets of depression.

Authors:  Rongrong Du; Qian Zhou; Tianzheng Hu; Jinmei Sun; Qiang Hua; Yingru Wang; Yuanyuan Zhang; Kongliang He; Yanghua Tian; Gong-Jun Ji; Kai Wang
Journal:  Front Psychol       Date:  2022-09-02

Review 5.  Optical neuroimaging: advancing transcranial magnetic stimulation treatments of psychiatric disorders.

Authors:  Shixie Jiang; Linda L Carpenter; Huabei Jiang
Journal:  Vis Comput Ind Biomed Art       Date:  2022-09-08

6.  Optimal coil orientation for transcranial magnetic stimulation.

Authors:  Lars Richter; Gunnar Neumann; Stephen Oung; Achim Schweikard; Peter Trillenberg
Journal:  PLoS One       Date:  2013-04-11       Impact factor: 3.240

Review 7.  Safety and recommendations for TMS use in healthy subjects and patient populations, with updates on training, ethical and regulatory issues: Expert Guidelines.

Authors:  Simone Rossi; Andrea Antal; Sven Bestmann; Marom Bikson; Carmen Brewer; Jürgen Brockmöller; Linda L Carpenter; Massimo Cincotta; Robert Chen; Jeff D Daskalakis; Vincenzo Di Lazzaro; Michael D Fox; Mark S George; Donald Gilbert; Vasilios K Kimiskidis; Giacomo Koch; Risto J Ilmoniemi; Jean Pascal Lefaucheur; Letizia Leocani; Sarah H Lisanby; Carlo Miniussi; Frank Padberg; Alvaro Pascual-Leone; Walter Paulus; Angel V Peterchev; Angelo Quartarone; Alexander Rotenberg; John Rothwell; Paolo M Rossini; Emiliano Santarnecchi; Mouhsin M Shafi; Hartwig R Siebner; Yoshikatzu Ugawa; Eric M Wassermann; Abraham Zangen; Ulf Ziemann; Mark Hallett
Journal:  Clin Neurophysiol       Date:  2020-10-24       Impact factor: 4.861

8.  Non-orthogonal one-step calibration method for robotized transcranial magnetic stimulation.

Authors:  He Wang; Jingna Jin; Xin Wang; Ying Li; Zhipeng Liu; Tao Yin
Journal:  Biomed Eng Online       Date:  2018-10-01       Impact factor: 2.819