Feasibility of the Engager™ aortic transcatheter valve system using a flexible over-the-wire design.

Simon H Sündermann1, Jürg Grünenfelder, Roberto Corti, Ardawan J Rastan, Axel Linke, Rüdiger Lange, Volkmar Falk, Sabine Bleiziffer.   

Abstract

OBJECTIVES: The aim was to investigate the safety and feasibility of the redesigned Engager™ transcatheter aortic valve implantation (TAVI) system.
METHODS: Transapical aortic valve implantation with the Engager™ valve prosthesis was intended in 11 patients, and performed in 10. Endpoints were defined according to the valve academic research consortium recommendations for reporting outcomes of TAVI in clinical trials.
RESULTS: All 10 patients were implanted successfully. No devicerelated or delivery system complications like coronary obstruction or aortic dissection emerged. One patient (10%) died from non-device-related reasons at post-operative day 23 of multi-organ failure. The invasively measured peak-to-peak gradient after valve implantation was 7.1 ± 3.5 mmHg. In 90%, there was no or only trivial (≤grad I) aortic regurgitation due to paravalvular leakage. In 10% of the patients, aortic regurgitation grade I-II was observed. At 30-day follow up, the mean gradient was 15.6 ± 4.9 mmHg, and no more than a mild transvalvular and paravalvular aortic regurgitation was seen as assessed by transthoracic echocardiography.
CONCLUSIONS: Application of the Engager™ TAVI system is safe and feasible. Prosthesis deployment in an anatomically correct position was facilitated by the design of the valve prosthesis and was successful in all patients. No device or delivery-system-related complications emerged. Safety and feasibility endpoints were met. Good results concerning the aortic valve performance after implantation and at 30-day follow up were ascertained. These results encouraged the start of a European Pivotal trial including patients to date.

Entities:  

Mesh:

Year:  2012        PMID: 22743081     DOI: 10.1093/ejcts/ezs389

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  3 in total

1.  Transapical aortic valve implantation with anatomically oriented prostheses.

Authors:  Simon Sündermann; Volkmar Falk
Journal:  Ann Cardiothorac Surg       Date:  2012-07

Review 2.  New devices for TAVI: technologies and initial clinical experiences.

Authors:  Maurizio Taramasso; Alberto Pozzoli; Azeem Latib; Giovanni La Canna; Antonio Colombo; Francesco Maisano; Ottavio Alfieri
Journal:  Nat Rev Cardiol       Date:  2014-01-21       Impact factor: 32.419

3.  Balloon- or Self-Expandable TAVI: Clinical Equipoise?

Authors:  John Jose; Gert Richardt; Mohamed Abdel-Wahab
Journal:  Interv Cardiol       Date:  2015-05
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.