Literature DB >> 22738842

A Monte Carlo simulation study of a flat-panel X-ray source.

Edwin J Grant1, Chrystian M Posada, Carlos H Castaño, Hyoung K Lee.   

Abstract

A novel flat-panel transmission X-ray source based on nitrogen-incorporated ultra-nano-crystalline diamond (N-UNCD) field emitters is being developed for medical and industrial X-ray imaging. X-ray generation characteristics of the X-ray tubes were simulated with the MCNPX 2.6.0 Monte Carlo code. The simulation results showed that 0.25-12 μm thick tungsten targets could generate an adequate intensity of Bremsstrahlung X-rays. Generated X-ray output intensities of 1.48×10(12)MeV/mA-s of 100 kVp electrons hitting the target were found before collimation.
Copyright © 2012 Elsevier Ltd. All rights reserved.

Entities:  

Year:  2012        PMID: 22738842     DOI: 10.1016/j.apradiso.2012.04.011

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  Design and optimization of thin-film tungsten (W)-diamond target for multi-pixel X-ray sources.

Authors:  Yuewen Tan; Qinghao Chen; Shuang Zhou; Erik A Henriksen; Tiezhi Zhang
Journal:  Med Phys       Date:  2022-05-31       Impact factor: 4.506

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.