Dongping Zeng1, Xiangguang Shen, Limin He, Huanzhong Ding, Youzhi Tang, Yongxue Sun, Binghu Fang, Zhenling Zeng.
Abstract
A rapid liquid chromatography tandem mass spectrometric method was developed for the simultaneous determination of mequindox and its five metabolites (2-isoethanol mequindox, 2-isoethanol 1-desoxymequindox, 1-desoxymequindox, 1,4-bisdesoxymequindox, and 2-isoethanol bisdesoxymequindox) in porcine muscle, liver, and kidney, fulfilling confirmation criteria with two transitions for each compound with acceptable relative ion intensities. The method involved acid hydrolysis, purification by solid-phase extraction, and subsequent analysis with liquid chromatography tandem mass spectrometry using electrospray ionization operated in positive polarity with a total run time of 15 min. The decision limit values of five analytes in porcine tissues ranged from 0.6 to 2.9 μg/kg, and the detection capability values ranged from 1.2 to 5.7 μg/kg. The results of the inter-day study, which was performed by fortifying porcine muscle (2, 4, and 8 μg/kg), liver, and kidney (10, 20, and 40 μg/kg) samples on three separate days, showed that the accuracy of the method for the various analytes ranged between 75.3 and 107.2% with relative standard deviation less than 12% for each analyte.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22733513 DOI: 10.1002/jssc.201100902
Source DB: PubMed Journal: J Sep Sci ISSN: 1615-9306 Impact factor: 3.645