Literature DB >> 22713916

Liquid-phase microextraction in a microfluidic-chip--high enrichment and sample clean-up from small sample volumes based on three-phase extraction.

María D Ramos Payán1, Henrik Jensen, Nickolaj Jacob Petersen, Steen Honoré Hansen, Stig Pedersen-Bjergaard.   

Abstract

In this work, a microfluidic-chip based system for liquid-phase microextraction (LPME-chip) was developed. Sample solutions were pumped into the LPME-chip with a micro-syringe pump at a flow rate of 3-4 μL min(-1). Inside the LPME chip, the sample was in direct contact with a supported liquid membrane (SLM) composed of 0.2 μL dodecyl acetate immobilized in the pores of a flat membrane of polypropylene (25 μm thickness). On the other side of the SLM, the acceptor phase was present. The acceptor phase was either pumped at 1 μL min(-1) during extraction or kept stagnant (stop-flow). Amitriptyline, methadone, haloperidol, loperamide, and pethidine were selected as model analytes, and they were extracted from alkaline sample solution, through the SLM, and into 10 mM HCl or 100mM HCOOH functioning as acceptor phase. Subsequently, the acceptor phase was either analyzed off-line by capillary electrophoresis for exact quantification, or on-line by UV detection or electrospray ionization mass spectrometry for time profiling of concentrations. The LPME-chip was found to be highly effective, and extraction efficiencies were in the range of 52-91%. When the flow of acceptor phase was turned off during extraction (stop-flow), analyte enrichment increased linearly with the extraction time. After 10 min as an example, amitriptyline was enriched by a factor of 42 from only 30 μL sample solution, and after 120 min amitriptyline was enriched by a factor of 500 from 320 μL sample solution. This suggested that the LPME-chip has great potentials for very efficient analyte enrichments from limited sample volumes in the future.
Copyright © 2012 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22713916     DOI: 10.1016/j.aca.2012.05.023

Source DB:  PubMed          Journal:  Anal Chim Acta        ISSN: 0003-2670            Impact factor:   6.558


  1 in total

1.  Microfluidic Chip-Based Induced Phase Separation Extraction as a Fast and Efficient Miniaturized Sample Preparation Method.

Authors:  Yao Shen; Bo Chen; Han Zuilhof; Teris A van Beek
Journal:  Molecules       Date:  2020-12-23       Impact factor: 4.411

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.