BACKGROUND: Monitoring of unfractionated heparin therapy by activated partial thromboplastin time using the ex vivo method for determining the heparin therapeutic range (HTR) is the standard of practice. Many extrinsic and intrinsic factors influence accuracy of the HTR. This study investigates the affect of instrumentation and laboratory site on the accuracy of the ex vivo HTR method.
METHODS: Patients on unfractionated heparin are used to determine the HTR by published guidelines. Various instruments and different laboratories are compared to investigate the affect of these variables on the HTR.
RESULTS: The HTR is the same when the same model of instrument or even different models (same methodology) are used in the same laboratory. However, a significant clinical difference is observed when different hospital laboratories using same lot and instrument model are compared.
CONCLUSIONS: The ex vivo method for HTR is the best protocol currently available. The HTR should be determined and averaged on all of the same method instruments used in the laboratory. However, determination of the HTR in different laboratories should not be shared.
© 2012 Blackwell Publishing Ltd.

Entities:  

Keywords:  APTT; anticoagulants; healthcare costs; heparin

Mesh:

Substances:

Year:  2012        PMID: 22712447     DOI: 10.1111/j.1751-553X.2012.01445.x

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  3 in total

Review 1.  Measurement of non-VKA oral anticoagulants versus classic ones: the appropriate use of hemostasis assays.

Authors:  Jonathan Douxfils; Anne Tamigniau; Bernard Chatelain; Catherine Goffinet; Jean-Michel Dogné; François Mullier
Journal:  Thromb J       Date:  2014-11-04

2.  Monitoring of Unfractionated Heparin Therapy in the Intensive Care Unit Using a Point-of-Care aPTT: A Comparative, Longitudinal Observational Study with Laboratory-Based aPTT and Anti-Xa Activity Measurement.

Authors:  Benjamin Lardinois; Michaël Hardy; Isabelle Michaux; Geoffrey Horlait; Thomas Rotens; Hugues Jacqmin; Sarah Lessire; Pierre Bulpa; Alain Dive; François Mullier
Journal:  J Clin Med       Date:  2022-02-28       Impact factor: 4.241

3.  Validation of high concentrated thrombin time assay for unfractionated heparin monitoring.

Authors:  Dollapak Apipongrat; Pornnapa Police; Rattapan Lamool; Punnee Butthep; Wittawat Chantkran
Journal:  J Clin Lab Anal       Date:  2022-09-13       Impact factor: 3.124

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.