Noninvasive CPAP with face mask: comparison among new air-entrainment masks and the Boussignac valve.

Giovanni Mistraletti1, Matteo Giacomini, Giovanni Sabbatini, Riccardo Pinciroli, Elena S Mantovani, Michele Umbrello, Debora Palmisano, Paolo Formenti, Anne L L Destrebecq, Gaetano Iapichino.   

Abstract

BACKGROUND: The performances of 2 noninvasive CPAP systems (high flow and low flow air-entrainment masks) were compared to the Boussignac valve in 3 different scenarios.
METHODS: Scenario 1: pneumatic lung simulator with a tachypnea pattern (tidal volume 800 mL at 40 breaths/min). Scenario 2: Ten healthy subjects studied during tidal breaths and tachypnea. Scenario 3: Twenty ICU subjects enrolled for a noninvasive CPAP session. Differences between set and effective CPAP level and F(IO(2)), as well as the lowest airway pressure and the pressure swing around the imposed CPAP level, were analyzed. The lowest airway pressure and swing were correlated to the pressure-time product (area of the airway pressure curve below the CPAP level) measured with the simulator. P(aO(2)) was a subject's further performance index.
RESULTS: Lung simulator: Boussignac F(IO(2)) was 0.54, even if supplied with pure oxygen. The air-entrainment masks had higher swing than the Boussignac (P = .007). Pressure-time product correlated better with pressure swing (Spearman correlation coefficient [ρ] = 0.97) than with lowest airway pressure (ρ = 0.92). In healthy subjects, the high-flow air-entrainment mask showed lower difference between set and effective F(IO(2)) (P < .001), and lowest airway pressure (P < .001), compared to the Boussignac valve. In all measurements the Boussignac valve showed higher than imposed CPAP level (P < .001). In ICU subjects the high-flow mask had lower swing than the Boussignac valve (P = .03) with similar P(aO(2)) increase.
CONCLUSIONS: High-flow air-entrainment mask showed the best performance in human subjects. During high flow demand, the Boussignac valve delivered lower than expected F(IO(2)) and showed higher dynamic hyper-pressurization than the air-entrainment masks.
© 2013 Daedalus Enterprises.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 22709435     DOI: 10.4187/respcare.01598

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  2 in total

1.  Effects of non-invasive respiratory supports on inspiratory effort in moderate-severe COVID-19 patients. A randomized physiological study.

Authors:  Gioacchino Schifino; Maria L Vega; Lara Pisani; Irene Prediletto; Vito Catalanotti; Vittoria Comellini; Ilaria Bassi; Maurizio Zompatori; Marco Vito Ranieri; Stefano Nava
Journal:  Eur J Intern Med       Date:  2022-04-22       Impact factor: 7.749

2.  Performance of EasyBreath Decathlon Snorkeling mask for delivering continuous positive airway pressure.

Authors:  Alberto Noto; Claudia Crimi; Andrea Cortegiani; Massimiliano Giardina; Filippo Benedetto; Pietro Princi; Annalisa Carlucci; Lorenzo Appendini; Cesare Gregoretti
Journal:  Sci Rep       Date:  2021-03-10       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.