Clinical diagnosis of pulp inflammation based on pulp oxygenation rates measured by pulse oximetry.

Frank C Setzer1, Simony Hidee Hamoy Kataoka, Fernanda Natrielli, Eudes Gondim-Junior, Celso Luiz Caldeira.   

Abstract

INTRODUCTION: The objective of this study was to investigate correlations between pulp oxygenation rates (%SpO(2)) and clinical diagnoses of reversible pulpitis (RP), irreversible pulpitis (IP), or pulp necrosis (PN).
METHODS: Sixty patients who presented with a tooth with endodontic pathology were grouped according to a clinical diagnosis of either RP (n = 20), IP (n = 20), or PN (n = 20). The clinical diagnosis was based on the patient's dental history, periapical radiographs, clinical inspection, and percussion and thermal sensitivity testing. Pulse oximetry (PO) was used to determine pulp oxygenation rates. For every patient, one additional endodontically treated tooth (negative control [NC], n = 60) and one additional healthy tooth with healthy pulp status (positive control [PC], n = 60) were evaluated. Analysis of variance, the Tukey HSD test, and the Student's t test were used for statistical analysis.
RESULTS: The mean %SpO(2) levels were as follows: RP: 87.4% (standard deviation [SD] ±2.46), IP: 83.1% (SD ±2.29), PN: 74.6% (SD ±1.96), PC: 92.2% (SD ±1.84), and NC: 0% (SD ±0.0). There were statistically significant differences between RP, IP, and PN compared with NC and PC and between RP, IP, and PN (all P ≤ .01).
CONCLUSIONS: The evaluation of pulp oxygenation rates by PO may be a useful tool to determine the different inflammatory stages of the pulp to aid in endodontic diagnosis.
Copyright © 2012 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22703647     DOI: 10.1016/j.joen.2012.03.027

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  5 in total

1.  Pulp sensibility tests responses in patients with anxiety and depression.

Authors:  Narges Farhad-Mollashahi; Mahboubeh-Firouzkouhi Moghadam; Seyed-Mohammad-Javad Aslani; Forugh Mollashahi
Journal:  J Clin Exp Dent       Date:  2022-05-01

2.  Reference values for pulp oxygen saturation as a diagnostic tool in endodontics: a systematic review and meta-analysis.

Authors:  Paula Lambert; Sergio Augusto Quevedo Miguens; Caroline Solda; Juliana Tomaz Sganzerla; Leandro Azambuja Reichert; Carlos Estrela; Fernando Branco Barletta
Journal:  Restor Dent Endod       Date:  2020-10-05

3.  Evaluation of the Pulp Oxygen Saturation Reading after Tooth Bleaching: A Randomized Clinical Trial.

Authors:  Dilma Helena Neves Henriques; Ana Maria Hecke Alves; Larissa Fernanda Pottmaier; Lucas da Fonseca Roberti Garcia; Eduardo Antunes Bortoluzzi; Cleonice Silveira Teixeira
Journal:  Int J Dent       Date:  2022-04-28

4.  Assessment of Pulp Oxygen Saturation Levels by Pulse Oximetry for Pulpal Diseases -A Diagnostic Study.

Authors:  Bander Anusha; Koppolu Madhusudhana; Suneel Kumar Chinni; Yelloji Paramesh
Journal:  J Clin Diagn Res       Date:  2017-09-01

5.  Relationship between the difference in electric pulp test values and the diagnostic type of pulpitis.

Authors:  Huachao Sui; Yangyang Lv; Mo Xiao; Liwen Zhou; Feng Qiao; Jinxin Zheng; Cuicui Sun; Jieni Fu; Yufan Chen; Yimeng Liu; Jie Zhou; Ligeng Wu
Journal:  BMC Oral Health       Date:  2021-07-10       Impact factor: 2.757

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.