Detection of crestal radiolucencies around dental implants: an in vitro experimental study.

Yigit Sirin1, Sinan Horasan, Duygu Yaman, Cansu Basegmez, Cem Tanyel, Ali Aral, Koray Guven.   

Abstract

PURPOSE: The aim of this study was to compare the diagnostic potentials and practical advantages of different imaging modalities in detecting bone defects around dental implants.
MATERIALS AND METHODS: Crestal bone defects with sequentially larger diameters were randomly prepared around 100 implants that were inserted in bovine bone blocks. Conventional periapical radiography (PR), direct digital radiography (DDR), panoramic radiography (PANO), cone-beam computed tomography (CBCT), and multislice computed tomography (MSCT) were performed for all specimens. The diagnostic accuracies of the devices, confidence of the answers, subjective image quality, defect visibility in planar orientations, and duration of diagnosis were analyzed based on the interpretations of 7 calibrated observers.
RESULTS: The agreement levels of intra- and interobserver scores were rated good. PR, DDR, and CBCT were mostly more accurate than PANO and MSCT (P < .05). Confidence levels were positively correlated with the defect size (ρ = 0.20, P < .01), and that of DDR was the highest (P < .05). The subjective image quality of PR and DDR was higher than that of CBCT, PANO, and MSCT (P < .05 for all comparisons). Axial-coronal-sagittal visibilities of the defects were higher for CBCT compared with MSCT (P < .05). The diagnostic time was shorter for DDR (P < .05) and longer for the tomographic systems (P < .05) than for the other devices.
CONCLUSIONS: DDR may provide a faster and more confident diagnostic option that is as accurate as PR in detecting peri-implant radiolucencies. CBCT has a comparable potential to these intraoral systems but with slower decision making and lower image quality, whereas PANO and MSCT become more reliable when bone defects have a diameter that is at least 1.5 mm larger than that of the implant.
Copyright © 2012 American Association of Oral and Maxillofacial Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22698290     DOI: 10.1016/j.joms.2012.02.024

Source DB:  PubMed          Journal:  J Oral Maxillofac Surg        ISSN: 0278-2391            Impact factor:   1.895


  6 in total

1.  Assessment of buccal marginal alveolar peri-implant and periodontal defects using a cone beam CT system with and without the application of metal artefact reduction mode.

Authors:  K Kamburoglu; E Kolsuz; S Murat; H Eren; S Yüksel; C S Paksoy
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

2.  Accuracy of CBCT images in the assessment of buccal marginal alveolar peri-implant defects: effect of field of view.

Authors:  K Kamburoğlu; S Murat; C Kılıç; S Yüksel; H Avsever; A Farman; W C Scarfe
Journal:  Dentomaxillofac Radiol       Date:  2014-03-20       Impact factor: 2.419

Review 3.  Etiology and Measurement of Peri-Implant Crestal Bone Loss (CBL).

Authors:  Adrien Naveau; Kouhei Shinmyouzu; Colman Moore; Limor Avivi-Arber; Jesse Jokerst; Sreenivas Koka
Journal:  J Clin Med       Date:  2019-02-01       Impact factor: 4.241

4.  Detection of peri-implant bone defects using cone-beam computed tomography and digital periapical radiography with parallel and oblique projection.

Authors:  Bardia Vadiati Saberi; Negar Khosravifard; Farnaz Ghandari; Arash Hadinezhad
Journal:  Imaging Sci Dent       Date:  2019-12-24

Review 5.  Diagnostic accuracy of imaging examinations for peri-implant bone defects around titanium and zirconium dioxide implants: A systematic review and meta-analysis.

Authors:  Mariana Murai Chagas; Solange Kobayashi-Velasco; Thais Gimenez; Marcelo Gusmão Paraiso Cavalcanti
Journal:  Imaging Sci Dent       Date:  2021-11-18

6.  Attitude in Radiographic Post-Operative Assessment of Dental Implants among Italian Dentists: A Cross-Sectional Survey.

Authors:  Bianca Di Murro; Piero Papi; Pier Carmine Passarelli; Antonio D'Addona; Giorgio Pompa
Journal:  Antibiotics (Basel)       Date:  2020-05-07
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.