A vibro-tactile display for clinical monitoring: real-time evaluation.

Maryam Dosani1, Kate Hunc, Guy A Dumont, Dustin Dunsmuir, Pierre Barralon, Stephan K W Schwarz, Joanne Lim, J Mark Ansermino.   

Abstract

BACKGROUND: Vibro-tactile displays use human skin to convey information from physiological monitors to anesthesiologists, providing cues about changes in the status of the patient. In this investigation, we evaluated, in a real-time clinical environment, the usability and wearability of a novel vibro-tactile display belt recently developed by our group, and determined its accuracy in identifying events when used by anesthesiologists.
METHODS: A prospective observational study design was used. During routine anesthesia, a standard physiological monitor was connected to a software tool that used algorithms to automatically identify changing trends in mean noninvasive arterial blood pressure, expired minute ventilation, peak airway pressure, and end-tidal carbon dioxide partial pressure. The software was wirelessly interfaced to a vibro-tactile belt worn by the anesthesiologist. Each physiological variable was mapped to 1 of 4 tactor locations within the belt. The direction (increase/decrease) and 2 levels of change (small/large) were encoded in the stimulation patterns. A training session was completed by each anesthesiologist. The system was activated in real-time during anesthesia alongside routine physiological monitors. When the algorithms detected changes in the patient, the belt vibrated at the appropriate location with the pattern corresponding to the level and direction of change. Using a touch screen monitor the anesthesiologist was to enter the vibro-tactile message by first identifying the variable, then identifying the level and direction of change. Usability and wearability questionnaires were to be completed. The percentage of correct identification of the physiological trend, the direction of change, and the level of change were primary outcome variables. The mean usability score and wearability results were secondary outcome variables. We hypothesized that anesthesiologists would correctly identify the events communicated to them through the vibro-tactile belt 90% of the time, and that anesthesiologists would find the vibro-tactile belt usable and wearable.
RESULTS: Seventeen anesthesiologists evaluated the display during 57 cases. The belt was operational for a mean (SD) duration of 75 (41) minutes per case. Seven cases were excluded from analysis because of technical failures. Eighty-one percent (confidence interval [CI], 77% to 84%) of all stimuli were decoded. The physiological trend, the direction of change, and the level of change were correctly identified for 97.7% (CI 96%-99%), 94.9% (CI 92%-97%), and 93.5% of these stimuli (CI, 91%-96%), respectively. Fourteen anesthesiologists completed the usability and wearability questionnaires. The mean usability score was 4.8 of a maximum usability score of 7.
CONCLUSIONS: Anesthesiologists found a vibro-tactile belt to be wearable and usable and could accurately decode vibro-tactile messages in a real-time clinical environment.

Entities:  

Mesh:

Year:  2012        PMID: 22696607     DOI: 10.1213/ANE.0b013e31825d638c

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  3 in total

Review 1.  Augmenting Critical Care Patient Monitoring Using Wearable Technology: Review of Usability and Human Factors.

Authors:  Evismar Andrade; Leo Quinlan; Richard Harte; Dara Byrne; Enda Fallon; Martina Kelly; Siobhan Casey; Frank Kirrane; Paul O'Connor; Denis O'Hora; Michael Scully; John Laffey; Patrick Pladys; Alain Beuchée; Gearoid ÓLaighin
Journal:  JMIR Hum Factors       Date:  2021-05-25

2.  The usability of ventilators: a comparative evaluation of use safety and user experience.

Authors:  Plinio P Morita; Peter B Weinstein; Christopher J Flewwelling; Carleene A Bañez; Tabitha A Chiu; Mario Iannuzzi; Aastha H Patel; Ashleigh P Shier; Joseph A Cafazzo
Journal:  Crit Care       Date:  2016-08-20       Impact factor: 9.097

3.  Avatar-based versus conventional vital sign display in a central monitor for monitoring multiple patients: a multicenter computer-based laboratory study.

Authors:  Olivier Garot; Julian Rössler; Juliane Pfarr; Michael T Ganter; Donat R Spahn; Christoph B Nöthiger; David W Tscholl
Journal:  BMC Med Inform Decis Mak       Date:  2020-02-10       Impact factor: 2.796

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.