Fragmented red blood cells automated measurement is a useful parameter to exclude schistocytes on the blood film.

J-F Lesesve1, V Asnafi1, F Braun1, G Zini1.   

Abstract

INTRODUCTION: The diagnosis of thrombotic microangiopathies (TMA) or disorders that may mimic their features remains difficult. Mechanical hemolytic anemia with the detection of shistocytes on the blood smear is a cornerstone finding to assess the diagnosis, but microscopic evaluation of shistocytes is still problematic with wide interobserver variations. Some of the latest generation automated blood cell counters (ABCC) propose an original quantitative approach of fragmented red cells (FRC), aiming to be equivalent to the microscopic count. This parameter has been poorly evaluated.
METHODS: To assess the predictive value (PV) of this test, we conducted studies comparing automated and microscopic counts of FRC/schistocytes, based on the analysis of thousands samples in four university hospitals and using the 2 ABCC currently available (Siemens ADVIA series, Sysmex XE-2100).
RESULTS: Reference range for FRC was <0.3% for the ADVIA and <0.5% for the XE-2100. The presence of FRC below a threshold determined at 1% (ADVIA and XE-2100) had a negative PV close to 100% to exclude the presence of schistocyte on the blood smear, but in relationship with a poor PV value.
CONCLUSIONS: Our study validated the utility of the immediately available FRC parameter on ABCC to exclude schistocytes and the diagnosis of TMA.
© 2012 Blackwell Publishing Ltd.

Entities:  

Keywords:  Fragmented red cells; automated hematology analysis; diagnostic test; schistocyte; thrombotic microangiopathy

Mesh:

Year:  2012        PMID: 22694255     DOI: 10.1111/j.1751-553X.2012.01434.x

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  2 in total

1.  The Clinical Significance of Schistocytes: A Prospective Evaluation of the International Council for Standardization in Hematology Schistocyte Guidelines.

Authors:  Elise Schapkaitz; Michael Halefom Mezgebe
Journal:  Turk J Haematol       Date:  2016-10-31       Impact factor: 1.831

2.  Comparison between optical microscopy and the Sysmex XN-3000 for schistocyte determination in patients suspected of having schistocytosis.

Authors:  Chattree Hantaweepant; Natthaporn Sasijareonrat; Boonyanuch Chutvanichkul; Khemajira Karaketklang; Yingyong Chinthammitr
Journal:  Health Sci Rep       Date:  2019-11-29
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.