A randomized controlled trial comparing the AccuVein AV300 device to standard insertion technique for intravenous cannulation of anesthetized children.

Roland N Kaddoum1, Doralina L Anghelescu, Mary E Parish, Becky B Wright, Luis Trujillo, Jianrong Wu, Yanan Wu, Laura L Burgoyne.   

Abstract

OBJECTIVES AND AIMS: To evaluate the efficacy of the AccuVein AV300 device in improving the first-time success rate of intravenous cannulation of anesthetized pediatric patients.
BACKGROUND: The AccuVein AV300 device was developed to assist venepuncture and intravenous cannulation by enhancing the visibility of superficial veins. It uses infrared light to highlight hemoglobin so that blood vessels are darkly delineated against a red background. METHODS/MATERIALS: Patients were randomized to cannulation with the AccuVein AV300 device or standard insertion by experienced pediatric anesthesiologists. An observer recorded the number of skin punctures and cannulation attempts required, and the time between tourniquet application and successful cannulation or four skin punctures, whichever came first. RESULTS AND
CONCLUSIONS: There were 146 patients with a median age of 4.6 years (range, 0.18-17.1 years), 46.6% were male, 80.8% were light skin colored, and 15.7% were younger than 2 years. The first-attempt success rates were 75% (95% CI, 63.8-84.2%) using AV300 and 73% (95% CI, 61.9-81.9%) using the standard method (P = 0.85). Patients with dark or medium skin color were 0.38 times less likely to have a successful first attempt than patients with light skin color. The difference between the two treatment groups in number of skin punctures and the time to insertion was not significant. Although the AV300 was easy to use and improved visualization of the veins, we found no evidence that it was superior to the standard method of intravenous cannulation in unselected pediatric patients under anesthesia.
© 2012 Blackwell Publishing Ltd.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22694242     DOI: 10.1111/j.1460-9592.2012.03896.x

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  9 in total

Review 1.  Utility of near-infrared light devices for pediatric peripheral intravenous cannulation: a systematic review and meta-analysis.

Authors:  Joon Min Park; Min Joung Kim; Hyeon Woo Yim; Won-Chul Lee; Hyunsuk Jeong; Na Jin Kim
Journal:  Eur J Pediatr       Date:  2016-10-26       Impact factor: 3.183

2.  A practical guiding method for vein graft harvesting.

Authors:  Salih Onur Basat; Tolga Aksan; Ufuk Askeroğlu; Mithat Akan
Journal:  J Hand Microsurg       Date:  2013-04-10

3.  Variations of the cubital superficial vein investigated by using the intravenous illuminator.

Authors:  Hyunsu Lee; Sang-Hoon Lee; Sung-Jin Kim; Woo-Ik Choi; Jae-Ho Lee; In-Jang Choi
Journal:  Anat Cell Biol       Date:  2015-03-20

4.  Multisensory System for the Detection and Localization of Peripheral Subcutaneous Veins.

Authors:  Roemi Fernández; Manuel Armada
Journal:  Sensors (Basel)       Date:  2017-04-19       Impact factor: 3.576

5.  Preliminary Study for Designing a Novel Vein-Visualizing Device.

Authors:  Donghoon Kim; Yujin Kim; Siyeop Yoon; Deukhee Lee
Journal:  Sensors (Basel)       Date:  2017-02-07       Impact factor: 3.576

6.  Initial experience with use of infrared assistance for intravenous injection of radiopharmaceuticals.

Authors:  Ashok Kumar; Meena Negi; Jyoti Khanka; Mohit Dhingra; Ranjeeta Kumari; Vandana Kumar Dhingra; Manoj Kumar Gupta
Journal:  World J Nucl Med       Date:  2020-10-02

7.  Towards a low-cost mobile subcutaneous vein detection solution using near-infrared spectroscopy.

Authors:  Simon Juric; Vojko Flis; Matjaz Debevc; Andreas Holzinger; Borut Zalik
Journal:  ScientificWorldJournal       Date:  2014-04-30

8.  An observational study comparing the prototype device with the existing device for the effective visualization of invisible veins in elderly patients in Japan.

Authors:  Keiko Kimori; Junko Sugama; Toshio Nakatani; Kazuya Nakayama; Tosiaki Miyati; Hiromi Sanada
Journal:  SAGE Open Med       Date:  2015-11-27

9.  Non-Invasive Imaging of Preoperative Mapping of Superficial Veins in Free Flap Breast Reconstruction.

Authors:  Michael W Chu; Jonathon R Sarik; Liza C Wu; Joseph M Serletti; Jonathan Bank
Journal:  Arch Plast Surg       Date:  2016-01-15
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.