Literature DB >> 22685949

Sterilization analysis of contaminated healing abutments and impression copings.

Vanessa Browne1, Michael Flewelling, Mark Wierenga, Alisa Wilson, Ray Aprecio, Paul Richardson, Nikola Angelov, Neal Johnson.   

Abstract

This study investigated sterilization of used implant impression copings and healing abutments. Components were analyzed after contamination with Enterococcus foecalis, followed by multiple rounds of sterilization by both steam autoclave and Chemiclave protocols. The authors' results demonstrated that used components showed sterility equal to new components without any visible distortion. These data suggest that component resterilization and reuse may be justified or at least considered in clinical practice. Also, implications for cost savings in the placement of implants are advanced.

Mesh:

Substances:

Year:  2012        PMID: 22685949

Source DB:  PubMed          Journal:  J Calif Dent Assoc        ISSN: 1043-2256


  1 in total

1.  Evaluation of residual contamination on reused healing abutments.

Authors:  Sezgi Cinel Sahin; Kadriye Ayca Dere
Journal:  Clin Oral Investig       Date:  2021-03-25       Impact factor: 3.573