Aquilla Turk1, Raymond D Turner, Satoshi Tateshima, David Fiorella, Kyung-Sool Jang, Imran Chaudry, Michael Kelly.
Abstract
INTRODUCTION: Treatment of wide-necked bifurcation aneurysms often poses procedural and long-term outcome challenges. The initial preclinical experience with the Pulsar Vascular Aneurysm Neck Reconstruction Device (PVANRD) in a canine bifurcation model is described.Entities:
Keywords: Aneurysm; CT; MRI; angioplasty; arteriovenous malformation; artery; balloon; coil; complication; device; flow diverter; navigation; reconstruction; spinal cord; spine; stent; stroke; subarachnoid; thrombectomy; thrombolysis; vein; wide-necked
Mesh:
Year: 2012 PMID: 22661600 PMCID: PMC3686258 DOI: 10.1136/neurintsurg-2012-010312
Source DB: PubMed Journal: J Neurointerv Surg ISSN: 1759-8478 Impact factor: 5.836
Figure 1Pulsar Vascular Aneurysm Neck Reconstruction Device.
Device performance, vessel and aneurysm dimensions, and coils placed
| Animal/device # | Vessel dimensions (parent vessel/left outflow/right outflow) | Aneurysm dimension/neck size | Coil sizes | Performance compared with Y-stent | Coil retention aneurysm | Microwire through ANRD (1=better, 2=same, 3=worse) | Microcatheter through ANRD (1=better, 2=same, 3=worse) | Device movement | Comments |
| 1/01 | 4 mm/3.3 mm/4.2 mm | 12.8×12.1 mm/5.5 mm | No coils | Better | N/A | 1 | 1 | Yes | Undersize device |
| 1/02 | 4 mm/3.3 mm/4.2 mm | 12.8×12.1 mm/5.5 mm | 15 mm×30 cm | Better | Yes | 1 | 1 | Yes | Undersize device |
| 2/01 | 4.3 mm/3.3 mm/3.6 mm | 11.2×10.0 mm/6.5 mm | No coils | Better | N/A | 1 | 1 | Yes | Undersize device |
| 2/02 | 4.3 mm/3.3 mm/3.6 mm | 11.2×10.0 mm/6.5 mm | 15 mm×30 cm; 24 mm×40 cm | Better | Yes | 1 | 1 | Yes | Undersize device |
| 3/01 | 3.8 mm/3.4 mm/3.7 mm | 8.1×7.9 mm/3.9 mm | No coils | Better | N/A | 1 | 1 | No | Stable device |
| 3/02 | 3.8 mm/3.4 mm/3.7 mm | 8.1×7.9 mm/3.9 mm | No coils | Better | N/A | 1 | 1 | No | Stable device |
| 4/01 | 3.3 mm/3.4 mm/2.4 mm | 10.5×6.6 mm/4.9 mm | No coils | Better | N/A | 1 | 1 | No | Stable device |
| 4/02 | 3.3 mm/3.4 mm/2.4 mm | 10.5×6.6 mm/4.9 mm | 9 mm×20 cm; 6 mm×15 cm; 5 mm×15 cm; 7 mm×15 cm; 6 mm×15 cm | Better | Yes | 1 | 1 | No | Stable device |
| 5/01 | 3.8 mm/3.6 mm/3.2 mm | 9.8×7.8 mm/4.4 mm | 10 mm×30 cm; 8 mm×20 cm; 8 mm×20 cm; 10 mm×30 cm | Better | Yes | 1 | 1 | No | Stable device |
| 6/01 | 4.2 mm/3.7 mm/3.5 mm | 10.8×9.8 mm/5.4 mm | 12 mm×30 cm; 9 mm×20 cm | Better | Yes | 1 | 1 | No | Stable device |
| 7/01 | 5.0 mm/3.9 mm/4.1 mm | 15.5×12.2 mm/5.8 mm | No coils | Better | N/A | 1 | 1 | No | Stable device |
| 8/01 | 3.7 mm/3.7 mm/3.4 mm | 10.4×7.8 mm/4.4 mm | 14 mm×30 cm; 13 mm×30 cm; 12 mm×20 cm; 13 mm×22 cm; 9 mm×20 cm; 6 mm×15 cm | Better | Yes | 1 | 1 | No | Stable device |
ANRD, Aneurysm Neck Reconstruction Device.
Figure 2(A) Native digital subtraction angiography (DSA) image showing device deployed across aneurysm neck and microcatheter in aneurysm. (B) Native DSA image showing placement of a 14 mm×30 cm and two 13 mm×30 cm coils with stable device position at aneurysm neck and no coil loop prolapse.