Dosimetric comparison between two MLC systems commonly used for stereotactic radiosurgery and radiotherapy: a Monte Carlo and experimental study.

K Asnaashari1, James C L Chow, Mostafa Heydarian.   

Abstract

In this work dosimetric parameters of two multi-leaf collimator (MLC) systems, namely the beam modulator (BM), which is the MLC commercial name for Elekta "Synergy S" linear accelerator and Radionics micro-MLC (MMLC), are compared using measurements and Monte Carlo simulations. Dosimetric parameters, such as percentage depth doses (PDDs), in-plane and cross-plane dose profiles, and penumbras for different depths and field sizes of the 6 MV photon beams were measured using ionization chamber and a water tank. The collimator leakages were measured using radiographic films. MMLC and BM were modeled using the EGSnrc-based BEAMnrc Monte Carlo code and above dosimetric parameters were calculated. The energy fluence spectra for the two MLCs were also determined using the BEAMnrc and BEAMDP. Dosimetric parameters of the two MLCs were similar, except for penumbras. Leaf-side and leaf-end 80-20% dose penumbras at 10 cm depth for a 10×10 cm(2) field size were 4.8 and 5.1mm for MMLC and 5.3 mm and 6.3 mm for BM, respectively. Both Radionics MMLC and Elekta BM can be used effectively based on their dosimetric characteristics for stereotactic radiosurgery and radiotherapy, although the former showed slightly sharper dose penumbra especially in the leaf-end direction.
Copyright © 2012 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Mesh:

Year:  2012        PMID: 22658764     DOI: 10.1016/j.ejmp.2012.05.001

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Evaluation of Dosimetric Effect of Millennium and High Definition Multi Leaf Collimator Using 6 MV Photon Beams.

Authors:  Aswathi Raj; D Khanna; Hridya Vt; Sathish Padmanabhan; P Mohandass
Journal:  Asian Pac J Cancer Prev       Date:  2021-12-01

2.  Rounded leaf end effect of multileaf collimator on penumbra width and radiation field offset: an analytical and numerical study.

Authors:  Dong Zhou; Hui Zhang; Peiqing Ye
Journal:  Radiol Oncol       Date:  2015-08-21       Impact factor: 2.991

3.  Comparison of stereotactic plans for brain tumors with two different multileaf collimating systems.

Authors:  Livia Marrazzo; Margherita Zani; Stefania Pallotta; Daniela Greto; Silvia Scoccianti; Cinzia Talamonti; Giampaolo Biti; Marta Bucciolini
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.