Cochlear implant device failure secondary to cholesterol granuloma-mediated cochlear erosion.

Ryan Eric Neilan1, Karen Pawlowski, Brandon Isaacson, Peter S Roland.   

Abstract

STUDY
DESIGN: Case report.
SETTING: A tertiary care pediatric hospital. PATIENT: A 7-year-old boy complained of retroauricular pain 3.5 years after cochlear implantation. Temporal bone computed tomographic scan revealed a soft tissue density filling the mastoid and middle ear space. There was extensive osseous erosion involving the cochlear promontory that resulted in lateral displacement electrode array into the middle ear space. INTERVENTION: Mastoidectomy with removal of cochlear implant.
RESULTS: Pathologic examination of the soft tissue filling the mastoid cavity was consistent with a cholesterol granuloma. The cultures of the mastoid cavity demonstrated no growth. Examination of the explanted device revealed a cell layer covering the length of the electrode. The stylet tract was filled with cells that grew out through the silicone component. No birefringent cholesterol crystals were detected in the tissue surrounding the electrode. Biofilms were not identified.
CONCLUSION: The authors describe a unique case of a cholesterol granuloma of the middle ear and mastoid in a cochlear implant recipient causing erosion of the otic capsule and displacement of the cochlear implant electrode array into the middle ear space.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22643442     DOI: 10.1097/MAO.0b013e3182544fff

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  4 in total

1.  The pattern and degree of capsular fibrous sheaths surrounding cochlear electrode arrays.

Authors:  Reuven Ishai; Barbara S Herrmann; Joseph B Nadol; Alicia M Quesnel
Journal:  Hear Res       Date:  2017-02-17       Impact factor: 3.208

2.  Intracochlear inflammatory response to cochlear implant electrodes in humans.

Authors:  Mohammad Seyyedi; Joseph B Nadol
Journal:  Otol Neurotol       Date:  2014-10       Impact factor: 2.311

3.  Foreign Body Response to Silicone in Cochlear Implant Electrodes in the Human.

Authors:  Jennifer T O'Malley; Barbara J Burgess; Donald Galler; Joseph B Nadol
Journal:  Otol Neurotol       Date:  2017-08       Impact factor: 2.311

4.  Cochlear Implant Electrode Migration due to Cholesterol Granuloma: Cues from a Case.

Authors:  Andrea Di Laora; Isabelle Mosnier; Andrea Ciorba; Stefano Pelucchi; Olivier Sterkers; Daniele Bernardeschi
Journal:  J Int Adv Otol       Date:  2019-12       Impact factor: 1.017

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.