Literature DB >> 22633857

Life prediction of different commercial dental implants as influence by uncertainties in their fatigue material properties and loading conditions.

M A Pérez1.   

Abstract

Probabilistic analyses allow the effect of uncertainty in system parameters to be determined. In the literature, many researchers have investigated static loading effects on dental implants. However, the intrinsic variability and uncertainty of most of the main problem parameters are not accounted for. The objective of this research was to apply a probabilistic computational approach to predict the fatigue life of three different commercial dental implants considering the variability and uncertainty in their fatigue material properties and loading conditions. For one of the commercial dental implants, the influence of its diameter in the fatigue life performance was also studied. This stochastic technique was based on the combination of a probabilistic finite element method (PFEM) and a cumulative damage approach known as B-model. After 6 million of loading cycles, local failure probabilities of 0.3, 0.4 and 0.91 were predicted for the Lifecore, Avinent and GMI implants, respectively (diameter of 3.75mm). The influence of the diameter for the GMI implant was studied and the results predicted a local failure probability of 0.91 and 0.1 for the 3.75mm and 5mm, respectively. In all cases the highest failure probability was located at the upper screw-threads. Therefore, the probabilistic methodology proposed herein may be a useful tool for performing a qualitative comparison between different commercial dental implants.
Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22633857     DOI: 10.1016/j.cmpb.2012.04.013

Source DB:  PubMed          Journal:  Comput Methods Programs Biomed        ISSN: 0169-2607            Impact factor:   5.428


  2 in total

1.  Biomechanical evaluations of the long-term stability of dental implant using finite element modeling method: a systematic review.

Authors:  Seyed Aref Hosseini-Faradonbeh; Hamid Reza Katoozian
Journal:  J Adv Prosthodont       Date:  2022-06-27       Impact factor: 1.989

2.  Finite Element Analysis of Stress in Bone and Abutment-Implant Interface under Static and Cyclic Loadings.

Authors:  Saeed Nokar; Hamid Jalali; Farideh Nozari; Mahnaz Arshad
Journal:  Front Dent       Date:  2020-09-07