PURPOSE: Conventional navigation systems for minimally invasive orthopedic surgery require a secondary monitor to display guidance information generated with CT or MRI images. Newer systems use augmented reality to project surgical plans into binocular glasses. These surgical procedures are often mentally challenging and cumbersome to perform.
METHOD: A comprehensive surgical navigation system for direct guidance while minimizing radiation exposure was designed and built. System accuracy was evaluated using in vitro needle insertion experiments. The fluoroscopic-based navigation technique is combined with an existing laser guidance technique. As a result, the combined system is capable of surgical planning using two or more X-ray images rather than CT or MRI scans. Guidance information is directly projected onto the patient using two laser beams and not via a secondary monitor.
RESULTS: We performed 15 in vitro needle insertion experiments as well as 6 phantom pedicle screw insertion experiments to validate navigation system accuracy. The planning accuracy of the system was found to be 2.32 mm and 2.28°, while its overall guidance accuracy was found to be 2.40 mm and 2.39°. System feasibility was demonstrated by successfully performing percutaneous pin insertion on phantoms.
CONCLUSION: Quantitative and qualitative evaluations of the fluorolaser navigation system show that it can support accurate guidance and intuitive surgical tool insertion procedures without preoperative 3D image volumes and registration processes.

Entities:  

Mesh:

Year:  2012        PMID: 22627882     DOI: 10.1007/s11548-012-0743-0

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  21 in total

1.  Virtual fluoroscopy: computer-assisted fluoroscopic navigation.

Authors:  K T Foley; D A Simon; Y R Rampersaud
Journal:  Spine (Phila Pa 1976)       Date:  2001-02-15       Impact factor: 3.468

2.  Evaluation of a portable image overlay projector for the visualisation of surgical navigation data: phantom studies.

Authors:  K Gavaghan; T Oliveira-Santos; M Peterhans; M Reyes; H Kim; S Anderegg; S Weber
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-10-21       Impact factor: 2.924

3.  Image overlay guidance for needle insertion in CT scanner.

Authors:  Gabor Fichtinger; Anton Deguet; Ken Masamune; Emese Balogh; Gregory S Fischer; Herve Mathieu; Russell H Taylor; S James Zinreich; Laura M Fayad
Journal:  IEEE Trans Biomed Eng       Date:  2005-08       Impact factor: 4.538

4.  Image-guided laser projection for port placement in minimally invasive surgery.

Authors:  Jonathan Marmurek; Chris Wedlake; Utsav Pardasani; Roy Eagleson; Terry Peters
Journal:  Stud Health Technol Inform       Date:  2006

5.  Box-Behnken design: an alternative for the optimization of analytical methods.

Authors:  S L C Ferreira; R E Bruns; H S Ferreira; G D Matos; J M David; G C Brandão; E G P da Silva; L A Portugal; P S dos Reis; A S Souza; W N L dos Santos
Journal:  Anal Chim Acta       Date:  2007-07-23       Impact factor: 6.558

6.  Augmented reality and image overlay navigation with OsiriX in laparoscopic and robotic surgery: not only a matter of fashion.

Authors:  Francesco Volonté; François Pugin; Pascal Bucher; Maki Sugimoto; Osman Ratib; Philippe Morel
Journal:  J Hepatobiliary Pancreat Sci       Date:  2011-07       Impact factor: 7.027

7.  Stereotactic laser-guided approach to distal middle cerebral artery aneurysms.

Authors:  J M Malik; T Kamiryo; J Goble; N F Kassell
Journal:  Acta Neurochir (Wien)       Date:  1995       Impact factor: 2.216

8.  Long bone fracture reduction using a fluoroscopy-based navigation system: a feasibility and accuracy study.

Authors:  Yoram A Weil; Meir Liebergall; Rami Mosheiff; David L Helfet; Andrew D Pearle
Journal:  Comput Aided Surg       Date:  2007-09

9.  Percutaneous placement of pedicle screws in the lumbar spine using a bone mounted miniature robotic system: first experiences and accuracy of screw placement.

Authors:  Ioannis Pechlivanis; George Kiriyanthan; Martin Engelhardt; Martin Scholz; Sebastian Lücke; Albrecht Harders; Kirsten Schmieder
Journal:  Spine (Phila Pa 1976)       Date:  2009-02-15       Impact factor: 3.468

10.  CT scanning: a major source of radiation exposure.

Authors:  Philip W Wiest; Julie A Locken; Philip H Heintz; Fred A Mettler
Journal:  Semin Ultrasound CT MR       Date:  2002-10       Impact factor: 1.875

View more
  4 in total

1.  Tracked ultrasound snapshots in percutaneous pedicle screw placement navigation: a feasibility study.

Authors:  Tamas Ungi; Eric Moult; Joseph H Schwab; Gabor Fichtinger
Journal:  Clin Orthop Relat Res       Date:  2013-08-17       Impact factor: 4.176

2.  Fluoroscopy-based laser guidance system for linear surgical tool insertion depth control.

Authors:  Takehito Doke; Jack T Liang; Shinya Onogi; Yoshikazu Nakajima
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-06-04       Impact factor: 2.924

3.  Image Overlay Surgery Based on Augmented Reality: A Systematic Review.

Authors:  Laura Pérez-Pachón; Matthieu Poyade; Terry Lowe; Flora Gröning
Journal:  Adv Exp Med Biol       Date:  2020       Impact factor: 2.622

Review 4.  Recent Development of Augmented Reality in Surgery: A Review.

Authors:  P Vávra; J Roman; P Zonča; P Ihnát; M Němec; J Kumar; N Habib; A El-Gendi
Journal:  J Healthc Eng       Date:  2017-08-21       Impact factor: 2.682

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.