Validation of micro-CT against the section method regarding the assessment of marginal leakage of sealants.

X Chen1, V M J I Cuijpers, M W Fan, J E Frencken.   

Abstract

BACKGROUND: The aim of this study was to validate the micro-CT and related software against the section method using the stereomicroscope for marginal leakage assessment along the sealant-enamel interface.
METHODS: Pits and fissures of the occlusal surface of 10 teeth were sealed with a resin-fissure sealant material without acid etching, thermocycled for 5000 cycles, immersed in 50% silver nitrate for three hours and scanned using micro-CT. Teeth were embedded in epoxy resin and cut in three sections. The middle section was subjected to micro-CT and stereomicroscopy. Images were taken from the left and right sides of the sealant-enamel interface at both the left and the right site of the section. Two experienced evaluators assessed marginal leakage.
RESULTS: Both assessment instruments observed no leakage in 37 out of the 40 images evaluated. Leakage at the sealant-enamel interface was observed in three stereomicroscopy images only. A fracture line in the sealant was seen on eight stereomicroscopy images and observed in only two micro-CT images.
CONCLUSIONS: The quality of the micro-CT and related software used in the present study does not qualify it to replace the section method as the gold standard for marginal leakage assessment at the sealant-enamel interface of permanent teeth.
© 2012 Australian Dental Association.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22624761     DOI: 10.1111/j.1834-7819.2012.01687.x

Source DB:  PubMed          Journal:  Aust Dent J        ISSN: 0045-0421            Impact factor:   2.291


  6 in total

1.  Micro-computerized tomographic analysis of premolars restored with oval and circular posts.

Authors:  Carlo Rengo; Gianrico Spagnuolo; Gianluca Ametrano; Jelena Juloski; Sandro Rengo; Marco Ferrari
Journal:  Clin Oral Investig       Date:  2013-04-21       Impact factor: 3.573

2.  Reliability of marginal microleakage assessment by visual and digital methods.

Authors:  Fábio Augusto de Santi Alvarenga; Camila Pinelli; Leonor de Castro Monteiro Loffredo
Journal:  Eur J Dent       Date:  2015 Jan-Mar

3.  In vitro Evaluation of the Marginal Fit and Internal Adaptation of Zirconia and Lithium Disilicate Single Crowns: Micro-CT Comparison Between Different Manufacturing Procedures.

Authors:  Francesco Riccitiello; Massimo Amato; Renato Leone; Gianrico Spagnuolo; Roberto Sorrentino
Journal:  Open Dent J       Date:  2018-02-22

4.  Tomographic Evaluation of the Internal Adaptation for Recent Calcium Silicate-Based Pulp Capping Materials in Primary Teeth.

Authors:  A A Al Tuwirqi; E A El Ashiry; A Y Alzahrani; N Bamashmous; T A Bakhsh
Journal:  Biomed Res Int       Date:  2021-05-08       Impact factor: 3.411

5.  Comparison of the Ability of Two Brands of CBCT with That of SEM to Detect the Marginal Leakage of Class V Composite Resin Restorations.

Authors:  Mitra Karbasi Kheir; Leili Khayam
Journal:  ScientificWorldJournal       Date:  2021-06-30

6.  Microbial microleakage assessment of class V cavities restored with different materials and techniques: A laboratory study.

Authors:  Hossein Nematollahi; Ali Bagherian; Kiarash Ghazvini; Habibollah Esmaily; Mina Azadegan Mehr
Journal:  Dent Res J (Isfahan)       Date:  2017 Sep-Oct