| Literature DB >> 2261936 |
Abstract
Acoustic inhomogeneities of body tissue may deflect diagnostic ultrasound waves as well as therapeutic shock waves. The influence of incorrect ultrasound targeting and shock wave application due to these deflection effects has been examined quantitatively by means of a two-dimensional physical model. The accuracy of in-line and out-of-line targeting has been determined. Calculations on 2 stone patients show that in-line deflection effects remain below 2 mm, while they reach values up to 6 mm for out-of-line systems. These results, confirmed by first clinical observations done by others, indicate that in-line ultrasound targeting is superior to out-of-line targeting.Entities:
Mesh:
Year: 1990 PMID: 2261936 DOI: 10.1159/000463913
Source DB: PubMed Journal: Eur Urol ISSN: 0302-2838 Impact factor: 20.096