Importance of diameter-to-length ratio in selecting dental implants: a methodological finite element study.

V Demenko1, I Linetskiy, K Nesvit, H Hubalkova, V Nesvit, A Shevchenko.   

Abstract

Implant dimensions greatly influence load transfer characteristics and the lifetime of a dental system. Excessive stresses at peri-implant area may result in bone failure. Finding the critical point at the implant-bone interface and evaluating the influence of implant diameter-to-length ratio on adjacent bone stresses makes it possible to select implant dimensions. For this, different cylindrical implants were numerically analysed using geometrical models generated from computed tomography images of mandible with osseointegrated implants. All materials were assumed to be linearly elastic and isotropic. Masticatory load was applied in its natural direction, oblique to occlusal plane. Maximum von Mises stresses were located around the implant neck at the critical point of its intersection with the plane of loading and were functions of implant diameter-to-length ratio. It was demonstrated that there exists a certain spectrum of diameter-to-length ratios, which will keep maximum bone stresses at a preset level chosen in accordance with patient's bone strength.

Entities:  

Keywords:  dental implant; diameter-to-length ratio; implant–bone interface; masticatory load; von Mises stress

Mesh:

Substances:

Year:  2012        PMID: 22616897     DOI: 10.1080/10255842.2012.688110

Source DB:  PubMed          Journal:  Comput Methods Biomech Biomed Engin        ISSN: 1025-5842            Impact factor:   1.763


  5 in total

1.  Bone quality effect on short implants in the edentulous mandible: a finite element study.

Authors:  Chaowei Liu; Yifeng Xing; Yan Li; Yanjun Lin; Jianghan Xu; Dong Wu
Journal:  BMC Oral Health       Date:  2022-04-26       Impact factor: 3.747

2.  In Vivo and In Vitro Analyses of Titanium-Hydroxyapatite Functionally Graded Material for Dental Implants.

Authors:  Xinhua Wang; Chengpeng Wan; Xiaoxia Feng; Fuyan Zhao; Huiming Wang
Journal:  Biomed Res Int       Date:  2021-04-30       Impact factor: 3.411

3.  Impact of implant diameter and length on stress distribution in osseointegrated implants: A 3D FEA study.

Authors:  R Eazhil; Siva Vadivel Swaminathan; Madhan Gunaseelan; G Vijay Kannan; Chandrapandian Alagesan
Journal:  J Int Soc Prev Community Dent       Date:  2016 Nov-Dec

4.  Assessing the Effect of Dental Implants Thread Design on Distribution of Stress in Impact Loadings Using Three Dimensional Finite Element Method.

Authors:  Zarei I; Khajehpour S; Sabouri A; Haghnegahdar Az; Jafari K
Journal:  J Dent Biomater       Date:  2016-06

5.  Influence of Implant Dimensions in the Resorbed and Bone Augmented Mandible: A Finite Element Study.

Authors:  Doriana Agop Forna; Norina Consuela Forna; Sînziana Anca Butnaru Moldoveanu
Journal:  Contemp Clin Dent       Date:  2020-12-20