Accuracy of implant placement with a stereolithographic surgical template.

Michele Cassetta1, Luigi V Stefanelli, Matteo Giansanti, Sabrina Calasso.   

Abstract

PURPOSE: The possible advantages of stereolithographic (SLA) surgical template use have not been adequately demonstrated, and studies of the accuracy of computer-aided implant placement are few. The purpose of this in vivo study was to evaluate the accuracy of a computer-designed (SLA) surgical guide by comparing the three-dimensional positions of planned and placed implants.
MATERIALS AND METHODS: One hundred sixteen implants were inserted in partially and completely edentulous patients using multiple SLA templates. Preoperative and postoperative computed tomographic images were compared. Four deviation parameters--global, angular, depth, and lateral--were defined and calculated between the planned and actual implant positions and analyzed statistically.
RESULTS: The mean global deviations between planned and actual implant positions at the coronal and apical ends were 1.47 mm and 1.83 mm, respectively; the mean angular deviation was 5.09 degrees. There were significant linear correlations at the implant level between coronal and angular deviations and between coronal and apical deviations.
CONCLUSIONS: The study highlighted deviations between the postoperative position and the preoperative plan at the coronal and apical portions of the implant, as well as in the angulation of the implant. Although the reported deviation values were extremely high, they do not appear to have resulted in important clinical complications. This suggests the necessity of always keeping a safety zone of at least 2 mm to avoid critical anatomical structures.

Entities:  

Mesh:

Year:  2012        PMID: 22616060

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  14 in total

1.  Implant Bed Preparation with an Erbium, Chromium Doped Yttrium Scandium Gallium Garnet (Er,Cr: YSGG) Laser Using Stereolithographic Surgical Guide.

Authors:  Gülin Seymen; Zeynep Turgut; Gizem Berk; Ayşen Bodur
Journal:  J Lasers Med Sci       Date:  2013

2.  An assessment of template-guided implant surgery in terms of accuracy and related factors.

Authors:  Jee-Ho Lee; Ji-Man Park; Soung-Min Kim; Myung-Joo Kim; Jong-Ho Lee; Myung-Jin Kim
Journal:  J Adv Prosthodont       Date:  2013-11-28       Impact factor: 1.904

Review 3.  Clinical applications and effectiveness of guided implant surgery: a critical review based on randomized controlled trials.

Authors:  Marco Colombo; Carlo Mangano; Eitan Mijiritsky; Mischa Krebs; Uli Hauschild; Thomas Fortin
Journal:  BMC Oral Health       Date:  2017-12-13       Impact factor: 2.757

4.  Accuracy of densitometry of two cone beam computed tomography equipment in comparison with computed tomography.

Authors:  Sina Haghanifar; Sajad Yousefi; Ehsan Moudi; Farida Abesi; Ali Bijani; Ali Akbar Moghadamnia; Mehrdad Nabahati
Journal:  Electron Physician       Date:  2017-05-25

5.  Implant-Supported PMMA Monolithic Full-Arch Rehabilitation with Surgical Computer-Planned Guide and Immediate Provisional: A Case Report with One Year Follow-Up.

Authors:  Vincenzo Luca Zizzari; Gianmarco Tacconelli
Journal:  Case Rep Dent       Date:  2018-04-01

Review 6.  Accuracy of Guided Surgery via Stereolithographic Mucosa-Supported Surgical Guide in Implant Surgery for Edentulous Patient: a Systematic Review.

Authors:  Cheongbeom Seo; Gintaras Juodzbalys
Journal:  J Oral Maxillofac Res       Date:  2018-03-31

7.  Alveolar ridge preservation and primary stability as influencing factors on the transfer accuracy of static guided implant placement: a prospective clinical trial.

Authors:  Sigmar Schnutenhaus; Liesa Brunken; Cornelia Edelmann; Jens Dreyhaupt; Heike Rudolph; Ralph G Luthardt
Journal:  BMC Oral Health       Date:  2020-06-29       Impact factor: 2.757

8.  Computer-assisted, template-guided immediate implant placement and loading in the mandible: a case report.

Authors:  Thomas Spielau; Uli Hauschild; Joannis Katsoulis
Journal:  BMC Oral Health       Date:  2019-04-11       Impact factor: 2.757

Review 9.  Trends in computer-aided manufacturing in prosthodontics: a review of the available streams.

Authors:  Jaafar Abduo; Karl Lyons; Mohammed Bennamoun
Journal:  Int J Dent       Date:  2014-04-08

10.  Accuracy of mucosa supported guided dental implant surgery.

Authors:  Andoni Jones
Journal:  Clin Case Rep       Date:  2018-09-23
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.