Analytical evaluation of Sysmex UF-1000i for flow cytometric analysis of peritoneal fluid.

Sabrina Buoro1, Rosanna Gustinetti, Paola Dominoni, Maria Grazia Alessio, Patrizia Filisetti, Maria Grazia Lucà, Stefano Fagiuoli, Cosimo Ottomano, Giuseppe Lippi.   

Abstract

OBJECTIVE: To evaluate analytical performance of Sysmex UF-1000i for peritoneal fluid analysis.
METHODS: Functional sensitivity, imprecision, linearity and comparison studies were performed on peritoneal fluids.
RESULTS: Total imprecision was 1.6-4.7%, functional sensitivity 27/μL for white blood cell (WBC) and 32/μL for total nucleated cell (TNC) count. Linearity was excellent up to 983 cell/μL, carry-over <0.2%, correlation with manual microscopy always greater than 0.992.
CONCLUSION: The instrument exhibited optimal performance at the conventional WBC diagnostic thresholds.
Copyright © 2012 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2012        PMID: 22609897     DOI: 10.1016/j.clinbiochem.2012.05.005

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  4 in total

1.  Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF-1000i.

Authors:  Sabrina Buoro; Sara Apassiti Esposito; MariaGrazia Alessio; Alberto Crippa; Cosimo Ottomano; Giuseppe Lippi
Journal:  J Clin Lab Anal       Date:  2015-08-24       Impact factor: 2.352

2.  Automated analysis for differentiating leukocytes in body fluids using the software "biological liquid application" on ADVIA2120/2120i hematology analyzer.

Authors:  Alessandra Falda; Paolo Doretto
Journal:  J Clin Lab Anal       Date:  2018-05-25       Impact factor: 2.352

3.  Evaluation of flow cytometry for the detection of bacteria in biological fluids.

Authors:  Elisa Rubio; Yuliya Zboromyrska; Jordi Bosch; Mariana J Fernandez-Pittol; Berta I Fidalgo; Assumpta Fasanella; Anna Mons; Angely Román; Climent Casals-Pascual; Jordi Vila
Journal:  PLoS One       Date:  2019-08-07       Impact factor: 3.240

4.  Evaluation of Flow Cytometry for Cell Count and Detection of Bacteria in Biological Fluids.

Authors:  N Dossou; I Gaubert; C Moriceau; E Cornet; S le Hello; D Malandain
Journal:  Microbiol Spectr       Date:  2022-02-23
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.