Optical performance comparison of deflectable laparoscopes for laparoendoscopic single-site surgery.

Zachariah G Goldsmith1, Gastón M Astroza, Agnes J Wang, W Neal Simmons, Muhammad W Iqbal, Michael E Lipkin, Glenn M Preminger, Michael N Ferrandino.   

Abstract

BACKGROUND AND
PURPOSE: Flexible working angles and fine optical visualization are major requisite factors in performing laparoendoscopic single-site (LESS) urologic procedures. Multiple mechanical design approaches have been used to develop deflectable laparoscopes for LESS procedures. We compared the optical characteristics of three such devices using a bench top approach to simulate LESS in straight and deflected positions.
MATERIALS AND METHODS: A 10-mm fixed-rod rotating lens device (Storz EndoCameleon) and two 5-mm articulating devices (Olympus EndoEye and Stryker IdealEye) were compared using standard industry testing protocols for image resolution (United States Air Force-1951 test target), distortion (multifrequency grid distortion target), and color reproducibility (Gretag Macbeth color checker).
RESULTS: The 10-mm fixed-rod rotating lens system demonstrated the highest image resolution (5.04 line pairs/mm), but also the highest distortion (22.8%). Among the 5-mm flexible articulating laparoscopes, resolution was superior with the Olympus EndoEye (4.00 line pairs/mm) compared with the Stryker IdealEye (3.17 line pairs/mm). Distortion (7.0%) and color reproduction (1.18) were superior with the IdealEye vs the EndoEye (18.8 %, 1.27). Laparoscope deflection resulted in attenuation of resolution by 11% with both articulating models, but not with the fixed rod system.
CONCLUSIONS: Definition of these optical characteristics may inform further development and selection of laparoscopic systems optimized for LESS surgery. A narrow but flexible camera can be crucial in the limited working space available during these procedures. Further investigation is warranted to determine if these objective findings translate into improved surgeon performance.

Mesh:

Year:  2012        PMID: 22591273     DOI: 10.1089/end.2012.0140

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  2 in total

Review 1.  Transumbilical single-incision laparoscopic distal pancreatectomy: primary experience and review of the English literature.

Authors:  Dianbo Yao; Shuodong Wu; Yu Tian; Ying Fan; Jing Kong; Yongnan Li
Journal:  World J Surg       Date:  2014-05       Impact factor: 3.352

2.  A miniature robotic steerable endoscope for maxillary sinus surgery called PliENT.

Authors:  Julie Legrand; Mouloud Ourak; Laura Van Gerven; Vincent Vander Poorten; Emmanuel Vander Poorten
Journal:  Sci Rep       Date:  2022-02-10       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.