Literature DB >> 22588597

Validity and reliability of preoperative templating in total hip arthroplasty using a digital templating system.

Ari Bertz1, Kari Indrekvam, Mohamed Ahmed, Erling Englund, Arkan S Sayed-Noor.   

Abstract

OBJECTIVES: To evaluate the validity, interobserver reliability, and intraobserver reproducibility of a digital templating system, the Mdesk™ in preoperative templating in cemented and reverse hybrid total hip arthroplasty (THA).
MATERIALS AND METHODS: Validity was evaluated by comparing the planned cup size, stem size, CCD angles, and neck length with the components used in 129 patients operated with cemented and reverse hybrid THA. The reliability was measured by comparing the templating results of two surgeons with each other (interobserver) and the results of two templatings carried out by first surgeon (intraobserver). The leg length discrepancy was measured before and after the operation to assess the templating ability to correct it.
RESULTS: The Mdesk™ system showed good validity (kappa value ranged from 0.64 to 0.96), especially when one size over and under the planned size were included. No difference between cemented and cementless stems was found. The interobserver reliability ranged from fair (kappa 0.23) to substantial (kappa 0.61) while the intraobserver reproducibility ranged from substantial (kappa 0.70) to excellent (kappa 0.82). Templating and intraoperative measures succeeded to restore the leg length.
CONCLUSIONS: The Mdesk™ system has comparable validity and reliability with other templating systems used in clinical practice. We recommend that the same surgeon who does the preoperative radiographic templating to also perform the operation. Further studies are required to evaluate the results of succeeded templating in the long run.

Entities:  

Mesh:

Year:  2012        PMID: 22588597     DOI: 10.1007/s00256-012-1431-4

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  22 in total

1.  Preoperative planning for total hip arthroplasty. Quantitating its utility and precision.

Authors:  J L Knight; R D Atwater
Journal:  J Arthroplasty       Date:  1992       Impact factor: 4.757

2.  Comparison of preoperative templating with postoperative assessment in cementless total hip arthroplasty.

Authors:  Kuen Tak Suh; Sang Jin Cheon; Dae Woong Kim
Journal:  Acta Orthop Scand       Date:  2004-02

Review 3.  Preoperative planning for primary total hip arthroplasty.

Authors:  Alejandro González Della Valle; Douglas E Padgett; Eduardo A Salvati
Journal:  J Am Acad Orthop Surg       Date:  2005-11       Impact factor: 3.020

4.  Digital templating in total hip arthroplasty.

Authors:  James V Bono
Journal:  J Bone Joint Surg Am       Date:  2004       Impact factor: 5.284

5.  The utility and precision of analogue and digital preoperative planning for total hip arthroplasty.

Authors:  Alejandro González Della Valle; Fernando Comba; Nicole Taveras; Eduardo A Salvati
Journal:  Int Orthop       Date:  2007-04-03       Impact factor: 3.075

6.  The precision and usefulness of preoperative planning for cemented and hybrid primary total hip arthroplasty.

Authors:  Alejandro González Della Valle; Gastón Slullitel; Francisco Piccaluga; Eduardo A Salvati
Journal:  J Arthroplasty       Date:  2005-01       Impact factor: 4.757

7.  Comparison of analog and digital preoperative planning in total hip and knee arthroplasties. A prospective study of 173 hips and 65 total knees.

Authors:  Bertram The; Ron L Diercks; Peter M A van Ooijen; Jim R van Horn
Journal:  Acta Orthop       Date:  2005-02       Impact factor: 3.717

8.  The utility of digital templating in Total Hip Arthroplasty with Crowe type II and III dysplastic hips.

Authors:  Xin Zhao; Zhen-An Zhu; Jie Zhao; Mao-Qiang Li; Gang Wang; De-Gang Yu; Bo Yu
Journal:  Int Orthop       Date:  2010-03-14       Impact factor: 3.075

9.  Determination of accuracy of preoperative templating of noncemented femoral prostheses.

Authors:  L W Carter; D O Stovall; T R Young
Journal:  J Arthroplasty       Date:  1995-08       Impact factor: 4.757

10.  A new digital preoperative planning method for total hip arthroplasties.

Authors:  Hendrikus J A Crooijmans; Armand M R P Laumen; Carola van Pul; Jan B A van Mourik
Journal:  Clin Orthop Relat Res       Date:  2008-09-10       Impact factor: 4.176

View more
  11 in total

1.  Radiographic templating of total hip arthroplasty for femoral neck fractures.

Authors:  Dror Lakstein; Ira Bachar; Ronen Debi; Omri Lubovsky; Ornit Cohen; Zachary Tan; Ehud Atoun
Journal:  Int Orthop       Date:  2016-06-07       Impact factor: 3.075

2.  Acetate templating on digital images is more accurate than computer-based templating for total hip arthroplasty.

Authors:  Robert Petretta; Jason Strelzow; Nicholas E Ohly; Peter Misur; Bassam A Masri
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

Review 3.  Computed tomography for preoperative planning in total hip arthroplasty: what radiologists need to know.

Authors:  Alexander Huppertz; Sebastian Radmer; Moritz Wagner; Torsten Roessler; Bernd Hamm; Martin Sparmann
Journal:  Skeletal Radiol       Date:  2014-03-13       Impact factor: 2.199

4.  Total Hip Arthroplasty Digital Templating: Size Predicting Ability and Interobserver Variability.

Authors:  Veronica Montiel; Santiago Troncoso; Andrés Valentí-Azcárate; Juan Ramón Valentí-Nin; Jose María Lamo-Espinosa
Journal:  Indian J Orthop       Date:  2020-08-03       Impact factor: 1.251

5.  Digital templating in total hip arthroplasty: Additional anteroposterior hip view increases the accuracy.

Authors:  Sophia K Stigler; Franz J Müller; Sebastian Pfaud; Michael Zellner; Bernd Füchtmeier
Journal:  World J Orthop       Date:  2017-01-18

6.  Accuracy and Reliability of Preoperative On-screen Templating Using Digital Radiographs for Total Hip Arthroplasty.

Authors:  Jong Ki Shin; Seung Min Son; Tae Woo Kim; Won Chul Shin; Jung Sub Lee; Kuen Tak Suh
Journal:  Hip Pelvis       Date:  2016-12-28

7.  The accuracy of digital templating in uncemented total hip arthroplasty.

Authors:  Lukas A Holzer; Georg Scholler; Stefan Wagner; Jörg Friesenbichler; Werner Maurer-Ertl; Andreas Leithner
Journal:  Arch Orthop Trauma Surg       Date:  2018-12-06       Impact factor: 3.067

Review 8.  The accuracy and reliability of preoperative digital 2D templating in prosthesis size prediction in uncemented versus cemented total hip arthroplasty: a systematic review and meta-analysis.

Authors:  Joshua B V Smith; Habeeb Bishi; Chao Wang; Vipin Asopa; Richard E Field; David H Sochart
Journal:  EFORT Open Rev       Date:  2021-11-19

9.  Use of Acetate Templates Over Digital Radiographs for Templating in Total Hip Arthroplasty: Technique and Its Validation.

Authors:  Tarun Goyal; Siddharth S Sethy; Sajid Ansari; Lakshmana Das; Souvik Paul
Journal:  Indian J Orthop       Date:  2020-11-03       Impact factor: 1.251

10.  Is analog preoperative planning still applicable?-comparison of accuracy of analog and computer preoperative planning methods in total hip arthroplasty.

Authors:  Julian Dutka; Slawomir Kiepura; Mateusz Bukowczan
Journal:  Ann Transl Med       Date:  2021-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.